EN – ENGLISH
19
Disinfection (continued)
Method 3:
Autoclave
• With device cleaned and disassembled, wrap components to keep clean.
• Time and temperature parameters vary widely according to the type of autoclave and packaging material. Follow
manufacturer’s instructions for loading and operating the autoclave.
• Make sure direct steam exposure to all surfaces is possible.
• Pre cycle vacuum (2psia / 14kPa) then steam disinfect at 132°C to 134°C for 4 minutes.
• Once device is removed from the autoclave, allow components to reach room temperature before reassembly,
handling, storage or use.
• Following the disinfection procedure, subject to routine cleaning (see Cleaning section).
5. Remove from the washing machine and ensure that all instruments are dried. If necessary, complete drying with a
sterile absorbent, lint-free towel.
6. Check all components to ensure they do not contain organic debris or residue of the cleaning agent. Repeat the
process if a soil is still visible. (For additionally validated cleaning cycles, refer to Appendix B)
Appendix A : Bleach Dilution
Bleach solutions produce chlorine gas. Prepare in a well-ventilated area.
NOTE: Hypochlorite potency decreases as bleach ages. Use only recently purchased bleach to prepare decontamination
solutions. Do not use pre-diluted bleach. Prepare a fresh batch for each application. Hypochlorite bleach mixing guidelines to
obtain 20,000 ppm solution:
Bleach
Preparation Instructions
4.5%
For 4 units bleach into 5 units of water
5%
For 2 units bleach into 3 units of water
6%
For 1 unit bleach into 2 units of water
NOTE: The use of highly alkaline agents may cause some components to tarnish and/or corrode. While this does not affect
the performance of the device we recommend that you use an acidic neutralizing agent immediately after each use as this
will help keep your device looking like new.
Appendix B: Additionally Validated Cleaning Cycles
Pre- Wash
Wash
Rinsing
Thermal
Disinfection
(Drying
Optional)
Temperature
Hold
Time
Detergent Treatment
Cold water from
building supply
(less than 110°or
43°C)/4min
93°C
1 or more
min.
Alkaline
(e.g. neodisher®
SeptoClean)
Hot water
from building
supply/2 min
Optional
90°C/2o
min.
50°C/7min
90°C
1 or more
min.
Neutral Enzymatic
(e.g. Enzol per
manufacturer's
instructions)
Hot water
from building
supply/ 2 min.
Optional
90°C/2o
min.
Содержание MAYFIELD 2 A3100
Страница 2: ...EN ENGLISH 2...
Страница 24: ...FR FRAN AIS 23...
Страница 25: ...FR FRAN AIS 24 Unit de base MAYFIELD 2 A3100 3101 3102 Mode d emploi...
Страница 26: ...FR FRAN AIS 25 25...
Страница 48: ...IT ITALIANO 47...
Страница 49: ...IT ITALIANO 48 Unit base MAYFIELD 2 A3100 3101 3102 Manuale di istruzioni...
Страница 50: ...IT ITALIANO 49 49...
Страница 72: ...DE DEUTSCH 71...
Страница 73: ...DE DEUTSCH 72 MAYFIELD 2 Basiseinheit A3100 3101 3102 Bedienungsanleitung...
Страница 74: ...DE DEUTSCH 73 73...
Страница 96: ...ES ESPA OL 95...
Страница 97: ...ES ESPA OL 96 Unidad base MAYFIELD 2 A3100 3101 3102 Manual de instrucciones...
Страница 98: ...ES ESPA OL 97 97...
Страница 120: ...NL NEDERLANDS 119...
Страница 121: ...NL NEDERLANDS 120 MAYFIELD 2 basisunit A3100 3101 3102 Instructiehandleiding...
Страница 122: ...NL NEDERLANDS 121 121...