INFICON
Calibration | 16
074-471-P1E HAPSITE ER Operating Manual
383 / 436
Additional calibration points can be added to the curve by following Step 10 through
Step 39. Click
OK
when the
A calibration point exists already will not be added
message is displayed.
35
Review the curves for each analyte by clicking the
Calibration
tab.
36
Click each analyte to display the corresponding calibration curve. The curve
should fit the data points. The menu provides four curve fit options:
Linear,
Linear, Forced through the Origin, Quadratic and Quadratic, Forced
through the Origin
.
The RSD of the curve varies depending upon the curve fit selected.
37
Verify that the
RSD of RF%
is acceptable. It is recommended that the
RSD of
RF%
is 30% or lower.
Содержание HAPSITE ER
Страница 1: ...Operating Manual HAPSITE ER 074 471 P1E ...
Страница 2: ...INFICON Two Technology Place East Syracuse NY 13057 9714 ...
Страница 12: ...1 Declaration of Conformity INFICON 12 436 074 471 P1E HAPSITE ER Operating Manual 1 Declaration of Conformity ...
Страница 13: ...INFICON Declaration of Conformity 1 074 471 P1E HAPSITE ER Operating Manual 13 436 ...
Страница 56: ...9 Operate the Instrument in Portable Mode INFICON 56 436 074 471 P1E HAPSITE ER Operating Manual ...
Страница 184: ...12 ER IQ Software INFICON 184 436 074 471 P1E HAPSITE ER Operating Manual ...
Страница 208: ...13 Data Review INFICON 208 436 074 471 P1E HAPSITE ER Operating Manual 4 Double click on the desired data file ...
Страница 226: ...13 Data Review INFICON 226 436 074 471 P1E HAPSITE ER Operating Manual 5 Click Save 6 Click Run ...
Страница 237: ...INFICON Data Review 13 074 471 P1E HAPSITE ER Operating Manual 237 436 2 Check the Search in NIST box ...
Страница 238: ...13 Data Review INFICON 238 436 074 471 P1E HAPSITE ER Operating Manual 3 Click Search ...
Страница 248: ...13 Data Review INFICON 248 436 074 471 P1E HAPSITE ER Operating Manual 2 The report is displayed ...
Страница 250: ...13 Data Review INFICON 250 436 074 471 P1E HAPSITE ER Operating Manual 5 The report opens in Excel ...
Страница 275: ...INFICON Data Review 13 074 471 P1E HAPSITE ER Operating Manual 275 436 6 The identification is displayed on the screen ...
Страница 323: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 323 436 ...
Страница 330: ...15 Method Editor INFICON 330 436 074 471 P1E HAPSITE ER Operating Manual 2 Select Next until the SIM page is displayed ...
Страница 333: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 333 436 5 Enter the name of the chemical of interest ...
Страница 349: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 349 436 The Analyte List is displayed ...
Страница 371: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 371 436 ð Click OK ...
Страница 381: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 381 436 ...
Страница 389: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 389 436 49 Click OK ...
Страница 390: ...16 Calibration INFICON 390 436 074 471 P1E HAPSITE ER Operating Manual 16 3 Definition of Terms in the Calibrate Window ...
Страница 435: ......
Страница 436: ......