9 | Operate the Instrument in Portable Mode
INFICON
78 / 436
074-471-P1E HAPSITE ER Operating Manual
• When a background has been established, sample directly over the point of
concern. Once the TIC begins to increase, slowly move the probe away from the
sample. If a compound has been identified, it is displayed on the screen.
Procedure
1
If an
Analyze
method is going to be run after
Survey
, verify that the
appropriate sample configuration (concentrator) is installed.
2
When powered on or taken out of
Extended Standby
, HAPSITE ER
automatically starts preparing an
ER Survey
and
Analyze
method if the probe
is attached. Refer to Operate the Instrument in Portable Mode [
51].
3
Verify that the desired
Analyze
method is listed under the
Survey
method. In
the figure below, the
Analyze
method is
ER_Air_Tri-Bed_PPB_Standard
.
4
A
SYSTEM READY
message is displayed with a prompt to press
Survey
or
Analyze
when HAPSITE ER is ready to sample.
Содержание HAPSITE ER
Страница 1: ...Operating Manual HAPSITE ER 074 471 P1E ...
Страница 2: ...INFICON Two Technology Place East Syracuse NY 13057 9714 ...
Страница 12: ...1 Declaration of Conformity INFICON 12 436 074 471 P1E HAPSITE ER Operating Manual 1 Declaration of Conformity ...
Страница 13: ...INFICON Declaration of Conformity 1 074 471 P1E HAPSITE ER Operating Manual 13 436 ...
Страница 56: ...9 Operate the Instrument in Portable Mode INFICON 56 436 074 471 P1E HAPSITE ER Operating Manual ...
Страница 184: ...12 ER IQ Software INFICON 184 436 074 471 P1E HAPSITE ER Operating Manual ...
Страница 208: ...13 Data Review INFICON 208 436 074 471 P1E HAPSITE ER Operating Manual 4 Double click on the desired data file ...
Страница 226: ...13 Data Review INFICON 226 436 074 471 P1E HAPSITE ER Operating Manual 5 Click Save 6 Click Run ...
Страница 237: ...INFICON Data Review 13 074 471 P1E HAPSITE ER Operating Manual 237 436 2 Check the Search in NIST box ...
Страница 238: ...13 Data Review INFICON 238 436 074 471 P1E HAPSITE ER Operating Manual 3 Click Search ...
Страница 248: ...13 Data Review INFICON 248 436 074 471 P1E HAPSITE ER Operating Manual 2 The report is displayed ...
Страница 250: ...13 Data Review INFICON 250 436 074 471 P1E HAPSITE ER Operating Manual 5 The report opens in Excel ...
Страница 275: ...INFICON Data Review 13 074 471 P1E HAPSITE ER Operating Manual 275 436 6 The identification is displayed on the screen ...
Страница 323: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 323 436 ...
Страница 330: ...15 Method Editor INFICON 330 436 074 471 P1E HAPSITE ER Operating Manual 2 Select Next until the SIM page is displayed ...
Страница 333: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 333 436 5 Enter the name of the chemical of interest ...
Страница 349: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 349 436 The Analyte List is displayed ...
Страница 371: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 371 436 ð Click OK ...
Страница 381: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 381 436 ...
Страница 389: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 389 436 49 Click OK ...
Страница 390: ...16 Calibration INFICON 390 436 074 471 P1E HAPSITE ER Operating Manual 16 3 Definition of Terms in the Calibrate Window ...
Страница 435: ......
Страница 436: ......