15 | Method Editor
INFICON
328 / 436
074-471-P1E HAPSITE ER Operating Manual
Name
Each scan set can be assigned a name for identification purposes. This
entry is optional.
One of the column entries listed above must be highlighted to enable editing of the
Mass
list for that specific
Scan Set
.
Mass
The mass fragments of each column are entered in this column.
Mass
Width
The width, in tenths of an amu, around the mass which the mass
spectrometer will scan. For example, a
Mass Width
of 0.6 will scan 0.3
amu on each side of the peak.
Extra
This sets the number of extra scans, from 0 - 10, for each mass. Extra
scans lower the detection limits by increasing the intensity within the
mass spectrometer. Extra scans should be used when scanning for
compounds with concentrations of ppb or lower.
Dwell
Dwell
is the amount of time the software will search each for the
selected mass. The dwell can be set from 100 µs - 5,000 µs. 400 µs is
recommended. Increasing the
Dwell
decreases the detection limit.
Lead In
Lead In
determines the number of points the mass spectrometer will
skip prior to scanning the desired mass peak. Best practice is to set the
Lead In
to at least a 1000 µs delay prior to collecting data. The delay is
based on
Lead In
multiplied by the
Dwell
.
The
Mass Width, Extra, Dwell
and
Lead In
values for a new entry are automatically
populated based upon the entry listed above.
To fill any column with the entry listed above, click in the desired cell and press Ctrl+D.
15.9.2 SIM for Survey
The SIM page for Survey mode provides the ability to create only one scan set.
Содержание HAPSITE ER
Страница 1: ...Operating Manual HAPSITE ER 074 471 P1E ...
Страница 2: ...INFICON Two Technology Place East Syracuse NY 13057 9714 ...
Страница 12: ...1 Declaration of Conformity INFICON 12 436 074 471 P1E HAPSITE ER Operating Manual 1 Declaration of Conformity ...
Страница 13: ...INFICON Declaration of Conformity 1 074 471 P1E HAPSITE ER Operating Manual 13 436 ...
Страница 56: ...9 Operate the Instrument in Portable Mode INFICON 56 436 074 471 P1E HAPSITE ER Operating Manual ...
Страница 184: ...12 ER IQ Software INFICON 184 436 074 471 P1E HAPSITE ER Operating Manual ...
Страница 208: ...13 Data Review INFICON 208 436 074 471 P1E HAPSITE ER Operating Manual 4 Double click on the desired data file ...
Страница 226: ...13 Data Review INFICON 226 436 074 471 P1E HAPSITE ER Operating Manual 5 Click Save 6 Click Run ...
Страница 237: ...INFICON Data Review 13 074 471 P1E HAPSITE ER Operating Manual 237 436 2 Check the Search in NIST box ...
Страница 238: ...13 Data Review INFICON 238 436 074 471 P1E HAPSITE ER Operating Manual 3 Click Search ...
Страница 248: ...13 Data Review INFICON 248 436 074 471 P1E HAPSITE ER Operating Manual 2 The report is displayed ...
Страница 250: ...13 Data Review INFICON 250 436 074 471 P1E HAPSITE ER Operating Manual 5 The report opens in Excel ...
Страница 275: ...INFICON Data Review 13 074 471 P1E HAPSITE ER Operating Manual 275 436 6 The identification is displayed on the screen ...
Страница 323: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 323 436 ...
Страница 330: ...15 Method Editor INFICON 330 436 074 471 P1E HAPSITE ER Operating Manual 2 Select Next until the SIM page is displayed ...
Страница 333: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 333 436 5 Enter the name of the chemical of interest ...
Страница 349: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 349 436 The Analyte List is displayed ...
Страница 371: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 371 436 ð Click OK ...
Страница 381: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 381 436 ...
Страница 389: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 389 436 49 Click OK ...
Страница 390: ...16 Calibration INFICON 390 436 074 471 P1E HAPSITE ER Operating Manual 16 3 Definition of Terms in the Calibrate Window ...
Страница 435: ......
Страница 436: ......