11 | Laptop Operation
INFICON
144 / 436
074-471-P1E HAPSITE ER Operating Manual
11.3.2 Tenax Concentrator
The Tenax concentrator is also used for analyzing samples with concentration levels
in the low part per million to high part per trillion range. The Tenax concentrator will
not effectively concentrate compounds with boiling points below 80°C, but may be
more effective at concentrating compounds with higher boiling points.
WARNING
The Tenax concentrator will not effectively concentrate compounds with
boiling points below 80°C.
Once installed, the concentrator must be cleaned before sampling begins.
ü
Verify that the concentrator is installed.
1
When powered on (refer to Operate the Instrument in Portable Mode [
51]) or
taken out of
Extended Standy
(refer to Extended Standby [
105]), HAPSITE
ER will automatically start preparing a concentrator method. If the method that
HAPSITE ER begins preparing is not the desired one, refer to Selecting a New
149].
2
Power on the laptop by pushing the
POWER
button. Open
ER IQ
Software by
double-clicking on the
ER IQ
icon.
3
HAPSITE ER will begin preparing to run the default concentrator method. It will
heat all necessary components, check pressures, and run an AutoTune, if
necessary. Progress of the heaters is indicated by a bar graph on the laptop
screen.
4
When the AutoTune is finished, the following message will be displayed. Click
OK
.
Содержание HAPSITE ER
Страница 1: ...Operating Manual HAPSITE ER 074 471 P1E ...
Страница 2: ...INFICON Two Technology Place East Syracuse NY 13057 9714 ...
Страница 12: ...1 Declaration of Conformity INFICON 12 436 074 471 P1E HAPSITE ER Operating Manual 1 Declaration of Conformity ...
Страница 13: ...INFICON Declaration of Conformity 1 074 471 P1E HAPSITE ER Operating Manual 13 436 ...
Страница 56: ...9 Operate the Instrument in Portable Mode INFICON 56 436 074 471 P1E HAPSITE ER Operating Manual ...
Страница 184: ...12 ER IQ Software INFICON 184 436 074 471 P1E HAPSITE ER Operating Manual ...
Страница 208: ...13 Data Review INFICON 208 436 074 471 P1E HAPSITE ER Operating Manual 4 Double click on the desired data file ...
Страница 226: ...13 Data Review INFICON 226 436 074 471 P1E HAPSITE ER Operating Manual 5 Click Save 6 Click Run ...
Страница 237: ...INFICON Data Review 13 074 471 P1E HAPSITE ER Operating Manual 237 436 2 Check the Search in NIST box ...
Страница 238: ...13 Data Review INFICON 238 436 074 471 P1E HAPSITE ER Operating Manual 3 Click Search ...
Страница 248: ...13 Data Review INFICON 248 436 074 471 P1E HAPSITE ER Operating Manual 2 The report is displayed ...
Страница 250: ...13 Data Review INFICON 250 436 074 471 P1E HAPSITE ER Operating Manual 5 The report opens in Excel ...
Страница 275: ...INFICON Data Review 13 074 471 P1E HAPSITE ER Operating Manual 275 436 6 The identification is displayed on the screen ...
Страница 323: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 323 436 ...
Страница 330: ...15 Method Editor INFICON 330 436 074 471 P1E HAPSITE ER Operating Manual 2 Select Next until the SIM page is displayed ...
Страница 333: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 333 436 5 Enter the name of the chemical of interest ...
Страница 349: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 349 436 The Analyte List is displayed ...
Страница 371: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 371 436 ð Click OK ...
Страница 381: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 381 436 ...
Страница 389: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 389 436 49 Click OK ...
Страница 390: ...16 Calibration INFICON 390 436 074 471 P1E HAPSITE ER Operating Manual 16 3 Definition of Terms in the Calibrate Window ...
Страница 435: ......
Страница 436: ......