Data Review | 13
074-471-P1E HAPSITE ER Operating Manual
259 / 436
Show Signal/Noise
Shows
Signal to Noise
ratio. A
background must be selected using
Background Subtract
first. Refer to
Subtract Current Spectrum
Subtracts the spectrum at the point where
the green scan cursor is located from the
range.
13.9 Displaying Reconstructed Ion Chromatograms
(RIC)
RIC plots are used to locate specific compounds in a chromatogram. A RIC plot of the
top three or more mass fragments can help locate the peak of interest. Follow the
instructions below to create a RIC plot.
The NIST program
uses the term peak
instead of mass
fragment. However,
the terms are
synonymous.
Alternately, double-clicking on a mass in the
Scan
window will automatically insert the
selected mass in the
Control Panel
table and display the RIC for the selected mass.
When the box in the
Control Panel
labeled
-RIC
is checked, the TIC/RIC window
displays the TIC minus the RIC selected.
1
Either from the
System Setup
screen or the
Data Review
screen, double-click
on the
NIST
icon.
2
Click on the
Names
tab at the bottom of the NIST screen.
3
Enter the name of the compound in the box on the top left of the screen. (i.e.,
benzene).
4
In the bottom right box, the
10 Largest Peaks
is listed. Make a note of the
three largest mass peaks that are between 45-300 amu.
Содержание HAPSITE ER
Страница 1: ...Operating Manual HAPSITE ER 074 471 P1E ...
Страница 2: ...INFICON Two Technology Place East Syracuse NY 13057 9714 ...
Страница 12: ...1 Declaration of Conformity INFICON 12 436 074 471 P1E HAPSITE ER Operating Manual 1 Declaration of Conformity ...
Страница 13: ...INFICON Declaration of Conformity 1 074 471 P1E HAPSITE ER Operating Manual 13 436 ...
Страница 56: ...9 Operate the Instrument in Portable Mode INFICON 56 436 074 471 P1E HAPSITE ER Operating Manual ...
Страница 184: ...12 ER IQ Software INFICON 184 436 074 471 P1E HAPSITE ER Operating Manual ...
Страница 208: ...13 Data Review INFICON 208 436 074 471 P1E HAPSITE ER Operating Manual 4 Double click on the desired data file ...
Страница 226: ...13 Data Review INFICON 226 436 074 471 P1E HAPSITE ER Operating Manual 5 Click Save 6 Click Run ...
Страница 237: ...INFICON Data Review 13 074 471 P1E HAPSITE ER Operating Manual 237 436 2 Check the Search in NIST box ...
Страница 238: ...13 Data Review INFICON 238 436 074 471 P1E HAPSITE ER Operating Manual 3 Click Search ...
Страница 248: ...13 Data Review INFICON 248 436 074 471 P1E HAPSITE ER Operating Manual 2 The report is displayed ...
Страница 250: ...13 Data Review INFICON 250 436 074 471 P1E HAPSITE ER Operating Manual 5 The report opens in Excel ...
Страница 275: ...INFICON Data Review 13 074 471 P1E HAPSITE ER Operating Manual 275 436 6 The identification is displayed on the screen ...
Страница 323: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 323 436 ...
Страница 330: ...15 Method Editor INFICON 330 436 074 471 P1E HAPSITE ER Operating Manual 2 Select Next until the SIM page is displayed ...
Страница 333: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 333 436 5 Enter the name of the chemical of interest ...
Страница 349: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 349 436 The Analyte List is displayed ...
Страница 371: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 371 436 ð Click OK ...
Страница 381: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 381 436 ...
Страница 389: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 389 436 49 Click OK ...
Страница 390: ...16 Calibration INFICON 390 436 074 471 P1E HAPSITE ER Operating Manual 16 3 Definition of Terms in the Calibrate Window ...
Страница 435: ......
Страница 436: ......