INFICON
Introduction | 7
074-471-P1E HAPSITE ER Operating Manual
27 / 436
The diaphragm pump consists of four stages. The diaphragm pump draws the gas
from the exhaust of the compound pump and sufficiently compresses exhaust gas in
order to discharge the exhaust into the atmosphere.
7.9 Electronic Systems
The electronic systems in HAPSITE ER are considered in four groups:
Mass Spectrometer
Control
The mass spectrometer control electronics include the programmable DC and RF
power supplies for the mass selector, DC power supplies for the filament, electron
multiplier, ion pump, and A/D converter for the signal from the electron multiplier.
Gas Chromatograph
Control
The Gas Chromatograph (GC) control circuitry includes the power supplies for the
solenoid valves, ovens and heated inlet line. It also controls the logic for all the valves
and heaters of the GC system.
Main Processor
The main processor is supported by solid state memory and is located in the central
electronics assembly. The main processor controls all the other electronic sub-
assemblies for routine operation.
Interfaces
There are several input/output devices within HAPSITE ER. These include the front
panel touchscreen, keypad and display, USB drive, crossover cable connection,
wireless connection, probe, power and logic connections to the Service Module, HSS,
SituProbe, TDSS, and SPME accessory.
7.10 Software Systems
HAPSITE ER operates with two software systems.
• Control software accepts inputs from the touchscreen, keypad and other
interfaces. It commands the operation and sequencing of all systems and
subsystems. The control software allows a method to be started from the front
panel. Design or modifications of a method require the use of ER IQ™ software on
an external laptop.
• Analysis software analyzes the data from the mass spectrometer, accesses the
libraries as required, and displays the results of the analyses on the front panel.
Additionally, HAPSITE ER application software, ER IQ, is a Windows XP, Windows
2000,Windows 7, and Windows 10 based system for laptop use. ER IQ is used to
design and modify methods, view data, analyze results, and generate reports. The
laptop is linked to HAPSITE ER via a specific crossover cable or wireless connection.
This linkage permits data and methods to be uploaded from HAPSITE ER. It also
allows for new or modified methods to be downloaded to HAPSITE ER.
The specific Windows version supported is based on the specific HAPSITE ER
hardware version and date of manufacture.
Содержание HAPSITE ER
Страница 1: ...Operating Manual HAPSITE ER 074 471 P1E ...
Страница 2: ...INFICON Two Technology Place East Syracuse NY 13057 9714 ...
Страница 12: ...1 Declaration of Conformity INFICON 12 436 074 471 P1E HAPSITE ER Operating Manual 1 Declaration of Conformity ...
Страница 13: ...INFICON Declaration of Conformity 1 074 471 P1E HAPSITE ER Operating Manual 13 436 ...
Страница 56: ...9 Operate the Instrument in Portable Mode INFICON 56 436 074 471 P1E HAPSITE ER Operating Manual ...
Страница 184: ...12 ER IQ Software INFICON 184 436 074 471 P1E HAPSITE ER Operating Manual ...
Страница 208: ...13 Data Review INFICON 208 436 074 471 P1E HAPSITE ER Operating Manual 4 Double click on the desired data file ...
Страница 226: ...13 Data Review INFICON 226 436 074 471 P1E HAPSITE ER Operating Manual 5 Click Save 6 Click Run ...
Страница 237: ...INFICON Data Review 13 074 471 P1E HAPSITE ER Operating Manual 237 436 2 Check the Search in NIST box ...
Страница 238: ...13 Data Review INFICON 238 436 074 471 P1E HAPSITE ER Operating Manual 3 Click Search ...
Страница 248: ...13 Data Review INFICON 248 436 074 471 P1E HAPSITE ER Operating Manual 2 The report is displayed ...
Страница 250: ...13 Data Review INFICON 250 436 074 471 P1E HAPSITE ER Operating Manual 5 The report opens in Excel ...
Страница 275: ...INFICON Data Review 13 074 471 P1E HAPSITE ER Operating Manual 275 436 6 The identification is displayed on the screen ...
Страница 323: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 323 436 ...
Страница 330: ...15 Method Editor INFICON 330 436 074 471 P1E HAPSITE ER Operating Manual 2 Select Next until the SIM page is displayed ...
Страница 333: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 333 436 5 Enter the name of the chemical of interest ...
Страница 349: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 349 436 The Analyte List is displayed ...
Страница 371: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 371 436 ð Click OK ...
Страница 381: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 381 436 ...
Страница 389: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 389 436 49 Click OK ...
Страница 390: ...16 Calibration INFICON 390 436 074 471 P1E HAPSITE ER Operating Manual 16 3 Definition of Terms in the Calibrate Window ...
Страница 435: ......
Страница 436: ......