14 | Tune
INFICON
284 / 436
074-471-P1E HAPSITE ER Operating Manual
Secondary Inputs/
Editables
• Focus Voltage
• Emission Current
• Ba Threshold
14.4.2 *Outputs
Base Peak Gain
(BPG)
BPG influences sensitivity.
• HAPSITE ER ideal setting is 0.5 (range is 0.4 to 0.6)
Ion Percentages
Actual Percentages are listed in the Mass Calibration table, outlined in red. The range
of acceptable values is presented in the 4 adjacent columns, outlined in blue in the
figure below.
Example:
The percentage of mass 50, first row, should fall within 0.5 to 2.5%, and
must fall within 0.4 to 3.0%, for the mass spec to be properly calibrated.
Boxes shaded with red, like those shown in the figure above, indicate that the actual
percentage is outside that range for that fragment (row). For example, in the third row
of the table above for mass 69, the actual percentage is 6.37 when it should fall above
8.00. Here, mass 69 requires adjustment of the Resolution and/or Ion Energy to
correct the Actual Percentage.
Status Column
The Status Column provides categorical view of the actual percentages. The values
OK, OK Low, OK High, Low
or
High
indicate whether the actual percentage for that
ion are within best range (OK) or outside (Low or High). Values of High or Low are not
acceptable. Adjust the corresponding mass fragments.
Содержание HAPSITE ER
Страница 1: ...Operating Manual HAPSITE ER 074 471 P1E ...
Страница 2: ...INFICON Two Technology Place East Syracuse NY 13057 9714 ...
Страница 12: ...1 Declaration of Conformity INFICON 12 436 074 471 P1E HAPSITE ER Operating Manual 1 Declaration of Conformity ...
Страница 13: ...INFICON Declaration of Conformity 1 074 471 P1E HAPSITE ER Operating Manual 13 436 ...
Страница 56: ...9 Operate the Instrument in Portable Mode INFICON 56 436 074 471 P1E HAPSITE ER Operating Manual ...
Страница 184: ...12 ER IQ Software INFICON 184 436 074 471 P1E HAPSITE ER Operating Manual ...
Страница 208: ...13 Data Review INFICON 208 436 074 471 P1E HAPSITE ER Operating Manual 4 Double click on the desired data file ...
Страница 226: ...13 Data Review INFICON 226 436 074 471 P1E HAPSITE ER Operating Manual 5 Click Save 6 Click Run ...
Страница 237: ...INFICON Data Review 13 074 471 P1E HAPSITE ER Operating Manual 237 436 2 Check the Search in NIST box ...
Страница 238: ...13 Data Review INFICON 238 436 074 471 P1E HAPSITE ER Operating Manual 3 Click Search ...
Страница 248: ...13 Data Review INFICON 248 436 074 471 P1E HAPSITE ER Operating Manual 2 The report is displayed ...
Страница 250: ...13 Data Review INFICON 250 436 074 471 P1E HAPSITE ER Operating Manual 5 The report opens in Excel ...
Страница 275: ...INFICON Data Review 13 074 471 P1E HAPSITE ER Operating Manual 275 436 6 The identification is displayed on the screen ...
Страница 323: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 323 436 ...
Страница 330: ...15 Method Editor INFICON 330 436 074 471 P1E HAPSITE ER Operating Manual 2 Select Next until the SIM page is displayed ...
Страница 333: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 333 436 5 Enter the name of the chemical of interest ...
Страница 349: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 349 436 The Analyte List is displayed ...
Страница 371: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 371 436 ð Click OK ...
Страница 381: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 381 436 ...
Страница 389: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 389 436 49 Click OK ...
Страница 390: ...16 Calibration INFICON 390 436 074 471 P1E HAPSITE ER Operating Manual 16 3 Definition of Terms in the Calibrate Window ...
Страница 435: ......
Страница 436: ......