INFICON
Method Editor | 15
074-471-P1E HAPSITE ER Operating Manual
337 / 436
There are four choices in the
Search Mode
drop down menu.
SIM Methods only
allow No Search as
the search option.
No Search
If this option is selected, a library search will not be conducted and
a report will not be displayed on the front panel at the end of the
run.
Qualitative
Searches AMDIS during a run to provide near real-time
identifications. A report will be generated at the end of the run and
can be viewed on the front panel display.
Quantitative
Generates a
Quantitative (Quant)
report at the end of the run by
referencing the designated library.
Qualitative/
Quantitative
Searches AMDIS to provide identifications during a run and
generates a
Quantitative
report.
Содержание HAPSITE ER
Страница 1: ...Operating Manual HAPSITE ER 074 471 P1E ...
Страница 2: ...INFICON Two Technology Place East Syracuse NY 13057 9714 ...
Страница 12: ...1 Declaration of Conformity INFICON 12 436 074 471 P1E HAPSITE ER Operating Manual 1 Declaration of Conformity ...
Страница 13: ...INFICON Declaration of Conformity 1 074 471 P1E HAPSITE ER Operating Manual 13 436 ...
Страница 56: ...9 Operate the Instrument in Portable Mode INFICON 56 436 074 471 P1E HAPSITE ER Operating Manual ...
Страница 184: ...12 ER IQ Software INFICON 184 436 074 471 P1E HAPSITE ER Operating Manual ...
Страница 208: ...13 Data Review INFICON 208 436 074 471 P1E HAPSITE ER Operating Manual 4 Double click on the desired data file ...
Страница 226: ...13 Data Review INFICON 226 436 074 471 P1E HAPSITE ER Operating Manual 5 Click Save 6 Click Run ...
Страница 237: ...INFICON Data Review 13 074 471 P1E HAPSITE ER Operating Manual 237 436 2 Check the Search in NIST box ...
Страница 238: ...13 Data Review INFICON 238 436 074 471 P1E HAPSITE ER Operating Manual 3 Click Search ...
Страница 248: ...13 Data Review INFICON 248 436 074 471 P1E HAPSITE ER Operating Manual 2 The report is displayed ...
Страница 250: ...13 Data Review INFICON 250 436 074 471 P1E HAPSITE ER Operating Manual 5 The report opens in Excel ...
Страница 275: ...INFICON Data Review 13 074 471 P1E HAPSITE ER Operating Manual 275 436 6 The identification is displayed on the screen ...
Страница 323: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 323 436 ...
Страница 330: ...15 Method Editor INFICON 330 436 074 471 P1E HAPSITE ER Operating Manual 2 Select Next until the SIM page is displayed ...
Страница 333: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 333 436 5 Enter the name of the chemical of interest ...
Страница 349: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 349 436 The Analyte List is displayed ...
Страница 371: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 371 436 ð Click OK ...
Страница 381: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 381 436 ...
Страница 389: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 389 436 49 Click OK ...
Страница 390: ...16 Calibration INFICON 390 436 074 471 P1E HAPSITE ER Operating Manual 16 3 Definition of Terms in the Calibrate Window ...
Страница 435: ......
Страница 436: ......