INFICON
Method Editor | 15
074-471-P1E HAPSITE ER Operating Manual
345 / 436
Window Expand Factor
This option multiplies the retention time of the peak by the
Window
Expand Factor
to give a period of time by which the search window will be
expanded. For example, if the peak retention time is 10 minutes and the
Window Expand Factor
is set to its default setting of 0.05, the 10 minute
retention time will be multiplied by the 0.05
Window Expand Factor
to
equal 30 seconds. Then, 30 seconds is added to the
Search Window
. If
the default value of the
Search Window
is 20 seconds, adding 30 seconds
from the
Window Expand Factor
to the
Search Window
would increase
the search range to 50 seconds.
Min. Width
This value is the minimum number of scans per peak, which designates the
area measurement for peak integration. Any peaks with fewer scans than
this value will be disregarded by the software. Decreasing this number will
result in the software accepting broader peaks.
Min. Fit
This compares the mass intensities of the compound to those saved in the
library. Reasonable values depend on the selectivity of the calibration, but
typically 0.5 to 0.9 is used. A higher
Min. Fit
number is more discriminative.
Peak Resolution (dx)
This number indicates the minimum number of scans between two peaks.
It is used to determine whether a peak should be considered a single peak
or if the peak should be split into two separate peaks.
Max Width
This value is the maximum number of scans per peak, which designates
the area measurement for peak integration. Any peaks with more scans
than this value will be disregarded by the software. Increasing this number
will result in the software accepting broader peaks.
Min. Purity
This compares the purity level of the detected peak to the mass peak in the
library. Reasonable values depend on the selectivity of the calibration, but
typically 0.5 to 0.9 is used. A higher
Min. Purity
number is more
discriminative.
Noise Level Mult
The peak intensity must be greater than the product of the
Noise Level
Mult
number multiplied by the baseline noise in order to be identified as an
analyte.
Precedence Level
Determines if the search uses the global parameters or compound-specific
parameters. Leaving this set to zero allows the use of specific search
parameters for individual compounds as discussed in Calibration [
370].
Min. Area
This number discriminates against low responses which are usually
attributed to noise rather than analyte detection. Increase this number to
10,000 or more if false positives are encountered.
Max Width
The peak must contain less scans than the setpoint number.
Low Mass
NIST will only use masses above this setpoint to make an identification.
High Mass
NIST will only use masses below this setpoint to make an identification.
Minimum Match Factor
The net fit in AMDIS must be above this number.
Содержание HAPSITE ER
Страница 1: ...Operating Manual HAPSITE ER 074 471 P1E ...
Страница 2: ...INFICON Two Technology Place East Syracuse NY 13057 9714 ...
Страница 12: ...1 Declaration of Conformity INFICON 12 436 074 471 P1E HAPSITE ER Operating Manual 1 Declaration of Conformity ...
Страница 13: ...INFICON Declaration of Conformity 1 074 471 P1E HAPSITE ER Operating Manual 13 436 ...
Страница 56: ...9 Operate the Instrument in Portable Mode INFICON 56 436 074 471 P1E HAPSITE ER Operating Manual ...
Страница 184: ...12 ER IQ Software INFICON 184 436 074 471 P1E HAPSITE ER Operating Manual ...
Страница 208: ...13 Data Review INFICON 208 436 074 471 P1E HAPSITE ER Operating Manual 4 Double click on the desired data file ...
Страница 226: ...13 Data Review INFICON 226 436 074 471 P1E HAPSITE ER Operating Manual 5 Click Save 6 Click Run ...
Страница 237: ...INFICON Data Review 13 074 471 P1E HAPSITE ER Operating Manual 237 436 2 Check the Search in NIST box ...
Страница 238: ...13 Data Review INFICON 238 436 074 471 P1E HAPSITE ER Operating Manual 3 Click Search ...
Страница 248: ...13 Data Review INFICON 248 436 074 471 P1E HAPSITE ER Operating Manual 2 The report is displayed ...
Страница 250: ...13 Data Review INFICON 250 436 074 471 P1E HAPSITE ER Operating Manual 5 The report opens in Excel ...
Страница 275: ...INFICON Data Review 13 074 471 P1E HAPSITE ER Operating Manual 275 436 6 The identification is displayed on the screen ...
Страница 323: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 323 436 ...
Страница 330: ...15 Method Editor INFICON 330 436 074 471 P1E HAPSITE ER Operating Manual 2 Select Next until the SIM page is displayed ...
Страница 333: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 333 436 5 Enter the name of the chemical of interest ...
Страница 349: ...INFICON Method Editor 15 074 471 P1E HAPSITE ER Operating Manual 349 436 The Analyte List is displayed ...
Страница 371: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 371 436 ð Click OK ...
Страница 381: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 381 436 ...
Страница 389: ...INFICON Calibration 16 074 471 P1E HAPSITE ER Operating Manual 389 436 49 Click OK ...
Страница 390: ...16 Calibration INFICON 390 436 074 471 P1E HAPSITE ER Operating Manual 16 3 Definition of Terms in the Calibrate Window ...
Страница 435: ......
Страница 436: ......