Troubleshooting
Problem
Cause
Correction
Pump stopped and cannot be
switched on again.
Thermal cutout activated due to motor
over-temperature.
• Manual reset is necessary. Switch off device
or isolate the equipment from mains. Wait
approx. five minutes before restarting the
device.
• Ensure pump fan is working and pump is
placed in an area with an ambient temp. of
less than or equal to 40°C
Ultimate vacuum cannot be
reached.
Fittings or hose connection have leakage.
• Ensure that fittings are tightened properly, use
PTFE thread seal tape at thread if necessary.
Check hose connection for any cracks. Follow
instructions for “
disassembling top cover
”
under “
Maintenance and Cleaning
” section
to check inner hoses for leaks.
Valves not seated properly, incorrect
orientation or damaged.
• Contact
IKA
service department.
Diaphragm not seated in bore properly or
damaged.
• Contact
IKA
service department.
Incorrect washer configuration assembled
during replacement of diaphragm.
• Contact
IKA
service department.
Liquid or solid parts inside the pump
chamber
• Use gas ballast valve to purge system of
condensates. Remove the vacuum hose to the
gas inlet and open the gas ballast valve. Turn
the pump on and run it at idle.
• If problem persists, contact
IKA
service
department.
Knocking noise during pump
operation.
Incorrect washer configuration assembled
during replacement of diaphragm.
• Contact
IKA
service department.
If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department
- Send the instrument for repair, including a short description of the fault.
Technical data
Rated operating voltage
VAC
230 ± 10% (EURO)
115 ± 10% (USA)
Operating voltage range
VAC
200 – 240 (EURO)
100 – 120 (USA)
Frequency
Hz
50 / 60
Power input at nominal voltage
W
180
Rated current
A
0,8
Ultimate vacuum (absolute)
mbar
<10
Ultimate vacuum (absolute) with gas balast
mbar
15
No-load speed
rpm
1380 / 50 Hz
1680 / 60 Hz
Inlet
Hose nozzle NW 10
Outlet
Hose nozzle NW 10
Max. permissible outlet pressure (absolute)
bar
1
Max. pressure difference between inlet and outlet
bar
1,1
Dimensions (W x D x H)
mm
245 x 320 x 155
Weight
Kg
11,5
Degree of protection
IP 54
Permissible ambient temperature storage/operation
°C
- 10 to + 60 / + 5 to + 40
Permissible humidity
%
80 (temperatures up to 30 °C)
Motor protection
Thermal cutout
Operation at a terrestrial altitude
m
max. 2000
Materials in contact with medium
Components
Wetted parts
Vacuum crater insert
PTFE Carbon reinforced
Vacuum crater
ETFE Carbon fibre-reinforced
Diaphragm clamping disc
ETFE Carbon fibre-reinforced
Valve
EFPM
Diaphragm
PTFE
Inlet/outlet/fittings
ETFE
Outlet
ETFE
Subject to technical changes!
Warranty
In accordance with
IKA
warranty conditions, the warranty
period is 24 months. For claims under the warranty please
contact your local dealer. You may also send the machine
direct to our factory, enclosing the delivery invoice and giv-
ing reasons for the claim. You will be liable for freight costs.
The warranty does not cover worn out parts, nor does it
apply to faults resulting from improper use, insufficient care
or maintenance not carried out in accordance with the in-
structions in this operating manual.
Содержание EW-28710-96
Страница 16: ...31 30 IKA IKA MVP 10 basic IKA MVP 10 basic IKA RV 8 RV 10 IKA MVP 10 basic IKA MVP 10 basic...
Страница 17: ...33 32 IKA IKA 4 A Fig 1 I 4 90 www ika com Decontamination Certificate IKA IKA www ika com IKA www ika com...
Страница 20: ...39 38 IKA IKA MVP 10 basic IKA MVP 10 basic IKA RV 8 RV10 IKA MVP 10 basic IKA MVP 10 basic...
Страница 21: ...41 40 IKA IKA IKA IKA 4 A Fig 1 I 4 90 www ika com IKA www ika com IKA www ika com...
Страница 34: ......