Display shows three lines:
After a few minutes of running time, measure
the conductivity of the permeate and/or raw/
mixed water with an external conductivity
meter and enter the measured value in the
corresponding parameter:
3 LIMIT OUT
»
Press
+
at the same time
»
Press ENTER to calibrate
»
Press ESC to exit without calibrating
4 LIMIT ROH-/MIX
»
Press
+
at the same time
»
Press ENTER to calibrate
»
Press ESC to exit without calibrating
11.6.2 Calibration of the probe zero
point
»
Remove the conductivity probe from
the flow assembly and dry it.
»
Select
2A CAL ZERO
in the main
menu
A conductivity value is now displayed in the
upper line.
»
Confirm the selection with
The display value may be > 0 µS, but should
remain < 50 µS, otherwise the controller
would not accept the calibration. In this case
the conductivity probe may be faulty and
must be replaced. The display shows the
message "ZERO ERROR cal.
»
repeat the calibration or quit the
parameter without saving with
2 KAL. PERMEAT
3 GRENZW.
PERMEAT
- - -
2 CAL PERMEATE
3 LIMIT OUT
2A ZERO CALIB
2B SLOPE CALIB
READING 3.12
µS
ZERO OK
Содержание WL-ROL 160-1200 Series
Страница 1: ...FlexLine Reverse osmosis plants MANUAL WLROL EN E 8881184 WLROL ENj WL ROL 160 1200...
Страница 62: ...Page 62...
Страница 65: ...Page 65 13 Declaration of Conformity...
Страница 66: ...Page 66...
Страница 72: ...Page 72 This page intentionally left blank...
Страница 73: ...SPA Page 73 This page intentionally left blank...