Hydro-Control VI Installation Guide HD0455 Rev 1.9.0
6.3
The Sensor Configuration Screen – Page 3
The Next button displays the third screen as shown in Figure 30.
Figure 30: The sensor configuration screen - Page 3
Factory Settings
This section allows the factory calibration of the sensor to be set. This is necessary if the sensor
is a Hydro-Probe Orbiter and the arm is changed or if the sensor is a Hydro-Mix and the ceramic
faceplate is replaced.
To set the factory calibration, make sure the sensor faceplate is in air and then press the ‘Read
Air’ button. After a short delay, the new air frequency and amplitude settings will be displayed
in the entry boxes. Next hold the sensor so that the faceplate is immersed in water as per the
instructions in the sensor user guide and press the ‘Read Water’ button.
An alternative method to the factory calibration is to use the Automatic Calibration function.
This can be used to simplify the factory calibration process. After the Automatic Calibration is
performed, it may be necessary to recalibrate the recipes.
To use the Automatic Calibration function, make sure that the sensor faceplate is in air and
press the AutoCal button. After a short delay the Hydro-Control will display an indication as to
whether the AutoCal was successful or not.
Orbiter Arms
This section allows the Arm Type to be set.
Temperature Compensation Coefficients
This section allows the temperature compensation parameters to be changed. It may be
necessary to change these settings when using a Hydro-Probe Orbiter and changing the Orbiter
Arm. An Engineering Note is included with relevant Orbiter Arms which details the settings that
must be entered. With certain Hydro-Mix sensors these coefficients are set for each sensor in
the factory and should not be changed.
Measurement Mode
This allows different measurement modes to be selected.
If any settings are changed then they should be downloaded to the sensor using the ‘Write To
Sensor’ button.
Содержание Hydro-Control VI
Страница 4: ...4 Hydro Control VI Installation Guide HD0455 Rev 1 9 0 ...
Страница 6: ...6 Hydro Control VI Installation Guide HD0455 Rev 1 9 0 ...
Страница 8: ...8 Hydro Control VI Installation Guide HD0455 Rev 1 9 0 ...
Страница 30: ...Electrical Installation Chapter 3 30 Hydro Control VI Installation Guide HD0455 Rev 1 9 0 ...
Страница 54: ...Commissioning Chapter 4 54 Hydro Control VI Installation Guide HD0455 Rev 1 9 0 ...
Страница 68: ...System Design Chapter 5 68 Hydro Control VI Installation Guide HD0455 Rev 1 9 0 ...
Страница 88: ...Remote Support Chapter 7 88 Hydro Control VI Installation Guide HD0455 Rev 1 9 0 ...
Страница 102: ...Document Cross Reference Appendix D 102 Hydro Control VI Installation Guide HD0455 Rev 1 9 0 ...