4-22
ELISYS QUATTRO
User Manual HUMAN
Performing a Test Run
4.5.1.5
Combined Report
4.5.1.6
Validation Criteria
4.5.1.7
Qualitative Results
4.5.1.8
Events
The Combined Report shows all patient IDs,
results, flags etc. in one table. It gives a short over-
view about all relevant data.
The samples are identified by the patient ID. The
individual wells are identified (alphanumerically),
evaluated, calculated and, if necessary, flagged.
The window Validation Criteria indicates if
the control values of the test meet the defaulted cri-
teria.
If the values of the control well are within the limits
specified by the formula in this field, the test is con-
sidered valid and can therefore be evaluated.
If one of the criterias is failed, the test will not be
evaluated.
The window Qualitative Results provides
information regarding the cut-off value of the test.
Three options are available:
• Pos: positive result
• Neg: negative result
• Eqv: result in the greygreygrey range.
The Events window shows information on impor-
tant actions performed by the analyzer and subse-
quent reactions by the user.
• Important actions in this context are warning and
error messages; they are displayed in red.
• Reactions by the user on these actions are dis-
played in green.
• User actions such as manual definition of rea-
gents are also displayed in red.
Содержание Elisys Quattro
Страница 1: ...HUMAN ELISYS QUATTRO User Manual Elisys Quattro User Manual Cat No 16300 1...
Страница 6: ...TOC 4 ELISYS QUATTRO User Manual HUMAN...
Страница 26: ...2 12 ELISYS QUATTRO User Manual HUMAN System Basics...
Страница 66: ...5 8 ELISYS QUATTRO User Manual HUMAN Loading Unloading Disposal...
Страница 69: ...Service and Maintenance ELISYS QUATTRO User Manual HUMAN 6 3 Pipettor...
Страница 87: ...Technical Data ELISYS QUATTRO User Manual HUMAN 8 5...