Loading - Unloading - Disposal
ELISYS QUATTRO
User Manual HUMAN
5-5
5.1.6
Load Test Plates
CAUTION: Laser radiation - do not stare into beam!
1.
When the software requests a plate the LD LED lights up
green
(on the
instrument front panel next to the right flap) and the Load Plate dialogue
appears.
2.
Open flap (very right). The plate transport unit is already in a position directly
next to the plate compartment.
3.
Place test plate in holding frame
(see chapter 2.2.4 on page 2-5)
and push it in,
overcoming a slight resistance.
Check if the plate has been inserted correctly and make sure no strip extends
beyond the edge of the frame.
4.
Place holding frame onto the plate transport unit such that both pins of the
holding frame on the right sit in the openings of the slide on the plate trans-
port unit
(see chapter 2.2.4 on page 2-5)
.
5.
Close flap.
6.
In the Load Plate dialogue, enter the plate ID and click OK (or Scan,
when using bar coded plates).
7.
The test plate is first transported into the photometer to check for an ade-
quate strip filling; then the plate is placed in the pipettor area.
8.
Repeat this process to load several plates in succession.
Содержание Elisys Quattro
Страница 1: ...HUMAN ELISYS QUATTRO User Manual Elisys Quattro User Manual Cat No 16300 1...
Страница 6: ...TOC 4 ELISYS QUATTRO User Manual HUMAN...
Страница 26: ...2 12 ELISYS QUATTRO User Manual HUMAN System Basics...
Страница 66: ...5 8 ELISYS QUATTRO User Manual HUMAN Loading Unloading Disposal...
Страница 69: ...Service and Maintenance ELISYS QUATTRO User Manual HUMAN 6 3 Pipettor...
Страница 87: ...Technical Data ELISYS QUATTRO User Manual HUMAN 8 5...