Performing a Test Run
4.2.2
Input of Bar Coded Patient Samples
The patient editor only accepts letters and numbers as patient ID. The use of
any spe-cial characters such as "%", "-", "$", "€","'", "|", "/", "\", ";", "," or "."
when creating a patient ID or receiving from the host computer is forbidden.
As an alternative, patients may be entered via host. In this case you may start
directly with creating a worklist
(see chapter 4.3.1 on page 4-5)
.
You will find further information to the tests in the manual of the respective
test.
1. Insert samples into the
ELISYS QUATTRO
with
the worklist closed.
The patient bar codes are read.
The dialogue box depicted to the right appears on
the screen, showing the identified bar codes.
2. From the respective list box (with all available
assays), select the respective assay by clicking
on the column header.
3. Then assign the samples by clicking on the field
in the respective assay column.
4. Click on the Close to accept the selection.
5. Proceed accordingly for each further sample
rack.