Error Messages
ELISYS QUATTRO
User Manual HUMAN
7-3
No disposable tips
left
Pipettor tips used up or not found.
Reload required tips at the position indi-
cated on the monitor.
No fluid detected for
%1
Insufficient volume for sample %1 or
reagent %1.
The status is documented in the active
event log. Confirm message with retry
or ignore. No further action is possible.
No respond to com-
mand '%1'
General software error. Communica-
tion between PC and analyzer is
interrupted.
Push the Retry button. If the message
comes up again, push the Ignore but-
ton, turn the PC off and start it again.
Answer the questions "....worklist ...?"
and "is the analyzer still running"
accordingly. The system starts again
and continues with the worklist in proc-
ess.
Plate ID verification
error
Plate bar code cannot be read.
Missing or faulty bar code, incorrectly
selected bar code type (see scanner
setup). Enter the plate ID manually.
Plate transport %1
motor home error
Plate transport motor cannot find
home position.
Select the menu item Utilities |
Selftest to initialise the system
again.
Plate transport %1
motor movement
error
Error in plate transport motor no.
%1.
Push the Retry button. If the error
recurs, push the Ignore button.
Plate transport %1
motor not homed
Plate transport motor cannot find
home position.
Select the menu item Utilities |
Selftest to initialise the system
again. If the error persists, call service.
Plate transport car-
rier error
Plate holder does not move far
enough into the instrument, so that
the light barrier is not interrupted.
Push the Retry button. If the error
recurs, push the Ignore button.
If this error recurs, the plate transport
has to be adjusted new. Call service.
Plate transport EEP-
ROM error
EEPROM error in plate transport.
Call service.
Plate transport error
Error in plate transport in case of
increased resistance.
Push the Retry button. If the error
recurs, try to locate the cause of the
resistance (blocking) and eliminate it. If
no mechanical blockage is evident, call
service.
Plate transport sen-
sor error
Error in plate transport sensor.
Call service.
Rack scanner focus-
ing error
The barcode scanner of the reagent
and sample racks cannot be
focused.
Manual allocation of racks or samples
and reagents is possible. Call service.
Rack scanner motor
error
Motor error in scanner of reagent
and sample rack. Scanner firmware
does not work correctly. Electrical or
mechanical problems of scanner.
Manual allocation of racks or samples
and reagents is possible. Call service.
Rack scanner not
detected
Scanner of reagent and sample
racks is not connected.
Manual allocation of racks or samples
and reagents is possible. Call service.
Error:
Cause:
Action:
Содержание Elisys Quattro
Страница 1: ...HUMAN ELISYS QUATTRO User Manual Elisys Quattro User Manual Cat No 16300 1...
Страница 6: ...TOC 4 ELISYS QUATTRO User Manual HUMAN...
Страница 26: ...2 12 ELISYS QUATTRO User Manual HUMAN System Basics...
Страница 66: ...5 8 ELISYS QUATTRO User Manual HUMAN Loading Unloading Disposal...
Страница 69: ...Service and Maintenance ELISYS QUATTRO User Manual HUMAN 6 3 Pipettor...
Страница 87: ...Technical Data ELISYS QUATTRO User Manual HUMAN 8 5...