7-8
ELISYS QUATTRO
User Manual HUMAN
Error Messages
Table 7-5:
Module 44 - Washer
Module 45
Incubator
Table 7-6:
Module 45 - Incubator
Module
ID:
Error
number:
Cause:
Action
44
41
Argument error.
Call service.
44
45
EEPROM error.
Call service.
44
48 nn
Motor not homed. nn shows the motor.
None. The software will reset the washer
and try again.
44
4D nn
Motor has become jammed during
movement. nn shows the motor.
Clear the obstruction. Call service.
44
50
Plate carrier sensor error.
Call service.
44
53 41
Washer waste bottle is full.
Empty the washer waste bottle. Call
service.
44
53 52
Reagent level is low.
Refill the reagent bottle. Call service.
44
52 41
Aspirate pump error.
Call service.
44
52 44
Dispense pump error.
Call service.
44
52 57
Waste pump error.
Call service.
44
56
Valve error.
Call service.
44
57
EEPROM error.
Call service.
Module
ID:
Error
number:
Cause:
Action
45
41
Argument error.
Call service.
45
48 nn
Heater error. nn shows the incubator
group number
Call service.
45
4B 43
Shake calibration failed.
Call service.
45
4B 52
Shake position frequency out of range.
Call service.
45
4B 54
Shake position edge timeout.
Call service.
45
4B 49
Shake override.
Call service.
45
50
Plate carrier sensor error.
Call service.
45
54 nn
Temperature sensor error. nn shows the
sensor number or name.
Call service.
45
57
EEPROM error.
Call service.
Содержание Elisys Quattro
Страница 1: ...HUMAN ELISYS QUATTRO User Manual Elisys Quattro User Manual Cat No 16300 1...
Страница 6: ...TOC 4 ELISYS QUATTRO User Manual HUMAN...
Страница 26: ...2 12 ELISYS QUATTRO User Manual HUMAN System Basics...
Страница 66: ...5 8 ELISYS QUATTRO User Manual HUMAN Loading Unloading Disposal...
Страница 69: ...Service and Maintenance ELISYS QUATTRO User Manual HUMAN 6 3 Pipettor...
Страница 87: ...Technical Data ELISYS QUATTRO User Manual HUMAN 8 5...