2-6
ELISYS QUATTRO
- User Manual HUMAN
System Basics
Figure 2-5:
ELISYS QUATTRO
- Bottom drawer pulled out
2.3.1
Wash Buffer Container and Waste Bottles
A maximum of 3 bottles (1 x 1 litre, 2 x 2 litres) can be used for various wash buffers;
another position has to be reserved for the cleaning fluid (distilled water) to clean the
washer head.
The connection fitting consists of 4 colour-coded connection pairs: one tubing and
one level sensor each per bottle
(Loading see chapter 5.1 on page 5-1)
.
Two waste bottles are available for the wash unit. One waste bottle contains the liquid
waste which is pumped to the waste container which is positioned below the instru-
ment. The second bottle serves as overflow protection.
2.3.2
Wash Unit for Test Plates with Window
The test plates are moved into the wash unit by the system. The wash process and
the height adjustment of the washer head can be checked via the window.
2.3.3
Incubator Unit
4
Test plate compartment with plate transport unit
5a
Wash buffer container (square) and waste bottles (round)
5b
Wash unit for test plates
(see chapter 2.3.2 on page 2-6)
6
Incubators
(see chapter 2.3.3 on page 2-6)
7
Photometer
Содержание Elisys Quattro
Страница 1: ...HUMAN ELISYS QUATTRO User Manual Elisys Quattro User Manual Cat No 16300 1...
Страница 6: ...TOC 4 ELISYS QUATTRO User Manual HUMAN...
Страница 26: ...2 12 ELISYS QUATTRO User Manual HUMAN System Basics...
Страница 66: ...5 8 ELISYS QUATTRO User Manual HUMAN Loading Unloading Disposal...
Страница 69: ...Service and Maintenance ELISYS QUATTRO User Manual HUMAN 6 3 Pipettor...
Страница 87: ...Technical Data ELISYS QUATTRO User Manual HUMAN 8 5...