4-12
ELISYS QUATTRO
User Manual HUMAN
Performing a Test Run
4.4.1.6
Job List
4.4.2
Loading Samples and Reagents
CAUTION: Laser radiation - do not stare into beam!
Allocate the vials only into suitable locations. An incorrect allocation may
cause errors in dispensing.
Samples having particulate matter, turbity, lipaemia, foam, or erythrocyte de-
bris may require clarification by filtration or centrifugation before testing.
The use of primary tubes with gel separator requires a careful check of the
sample volume above the gel to avoid any possible errors in dispensing.
When refilling a reagent bottle never exceed the bottle shoulder. An over-fill-
ing may lead to an incorrect dispensing.
The Job list shows the allocation of patient IDs
and tests to be performed in a list.
Содержание Elisys Quattro
Страница 1: ...HUMAN ELISYS QUATTRO User Manual Elisys Quattro User Manual Cat No 16300 1...
Страница 6: ...TOC 4 ELISYS QUATTRO User Manual HUMAN...
Страница 26: ...2 12 ELISYS QUATTRO User Manual HUMAN System Basics...
Страница 66: ...5 8 ELISYS QUATTRO User Manual HUMAN Loading Unloading Disposal...
Страница 69: ...Service and Maintenance ELISYS QUATTRO User Manual HUMAN 6 3 Pipettor...
Страница 87: ...Technical Data ELISYS QUATTRO User Manual HUMAN 8 5...