Error Messages
ELISYS QUATTRO
User Manual HUMAN
7-7
Table 7-3:
Module 42 - Plate transport
Module 43
Photometer
Table 7-4:
Module 43 - Photometer
Module 44
Washer
42
4D nn
Motor jammed during movement. nn
shows the axis.
Call service.
42
50
Plate carrier sensor failure.
Call service.
42
52
Firmware was unable to compute a safe
route between the two positions.
Call service.
42
57
EEPROM error.
Call service.
Module
ID:
Error
number:
Cause:
Action
Module
ID:
Error
number:
Cause:
Action
43
41
Argument error in command.
Check that the firmware is the correct
version for the version of software being
run. Call service.
43
42
Background light levels too high.
Check position of lamp. Check filter posi-
tions. Call service.
43
43
A/D conversion error.
Call service.
43
46 nn
Filter error. nn shows the filter position.
Call service.
43
48 nn
Motor not homed. nn shows the motor.
None. The software will reset the pho-
tometer and try again.
43
4C
Lamp has blown.
Replace the lamp.
43
4D nn
Motor is jammed. nn shows the motor.
Call service.
43
50
Plate carrier error.
Check that the plate carrier correctly
breaks the sensor. Call service.
43
50 44
No scan data available.
Call service.
43
50 4F
A/D over-range error.
Call service.
43
50 55
A/D under-range error.
Call service.
43
50 57
Reading overrun error.
Call service.
43
52
A/D reference voltage error.
Call service.
43
54 nn
Faulty optic channel. nn shows the filter
position.
Call service.
43
57
EEPROM error.
Call service.
Содержание Elisys Quattro
Страница 1: ...HUMAN ELISYS QUATTRO User Manual Elisys Quattro User Manual Cat No 16300 1...
Страница 6: ...TOC 4 ELISYS QUATTRO User Manual HUMAN...
Страница 26: ...2 12 ELISYS QUATTRO User Manual HUMAN System Basics...
Страница 66: ...5 8 ELISYS QUATTRO User Manual HUMAN Loading Unloading Disposal...
Страница 69: ...Service and Maintenance ELISYS QUATTRO User Manual HUMAN 6 3 Pipettor...
Страница 87: ...Technical Data ELISYS QUATTRO User Manual HUMAN 8 5...