14 In case of
negative
result of the
M
N-PE
, the auto
test is automatically blocked, the
symbol is shown
and the screen to the side is displayed. Note red
value (lower than minimum set trhreshold) of the N-PE
insulation resistance
Press the
SAVE
button or touch the
icon to save
the measurement (see § 7.1)
AutoTest in TN systems
6.10.2.
6. Press the
GO/STOP
key
for few seconds
or the
START
key on the remote lead. During this whole
stage, do not disconnect the measuring leads of the
instrument from the system under test. In case of
RCD
tripping time measurement
selection the screen to
the side appears on the display when the hourglass
icon indicates the performance of the test
7. Press the
GO/STOP
key
for few seconds
or the
START
key on the remote lead. During this whole
stage, do not disconnect the measuring leads of the
instrument from the system under test. In case of
RCD
tripping current measurement
selection the screen to
the side appears on the display when the hourglass
icon indicates the performance of the test
8. In case of
positive
result of the three test sequentially
performed as
NoTrip
(
Z
L-N
and
Z
L-PE
<199
),
RCDX1
(see § 13.4) and
M
L/N-PE
the
symbol is shown
and the screen to the side is displayed by the
instrument
Press the
SAVE
button or touch the
icon to save
the measurement (see § 7.1)
9. In case of
positive
result of the three test sequentially
performed as
NoTrip
(
Z
L-N
and
Z
L-PE
<199
),
RCD
(see § 13.4) and
M
L/N-PE
the
symbol is shown
and the screen to the side is displayed by the
instrument
Press the
SAVE
button or touch the
icon to save
the measurement (see § 7.1)