Gx
Series
EN - 61
5. Press the
GO/STOP
key
for few seconds
or the
START
key on the remote lead. During this whole
stage, do not disconnect the measuring leads of the
instrument from the system under test.
In case of positive result (overall earth resistance lower
than the ratio between limit contact voltage and RCD
tripping current), the screen to the side is displayed by
the instrument, which contains the contact voltage
value in the secondary display.
Press the
SAVE
button or touch the
icon to save
the measurement (see § 7.1).
6. In case of negative result (overall earth resistance
higher than the ratio between limit contact voltage and
RCD tripping current), the screen to the side is
displayed by the instrument.
Note the presence of the measurement result of the
contact voltage highlighted in red.
Press the
SAVE
button or touch the
icon to save
the measurement (see § 7.1).