Gx
Series
EN - 22
AUTO L/N-PE mode
6.2.2.
1.
Touch the
icon. The screen to the side appears
on the display. The instrument automatically carries out
the test for the presence of voltage between the inputs
(shown on the display) and blocks the test in case of
voltage higher than 10V
Touch the “AUTO L/N-PE” icon to set the measuring
mode. The following screen appears on the display:
2. Move the slide bar reference in the positions "AUTO"
(Automatic mode) or " " (Timer mode). Move the right
slide bar reference in the “L/N-PE” position. Confirm the
choice by going back to the previous screen.
If Timer mode is selected, the following screen is
shown:
3. Touch the icon "R
xx
" to set the minimum limit value
of the insulation resistance on which the instrument
makes the comparison with the measured value. The
screen to the side appears on the display.
Touch the
icon to zero the value in the “R
” field.
Use the virtual keyboard to set the value between
0.01M
and
999M
. Confirm the choice by going back
to the initial measurement screen. Note the presence of
the set limit value.
4. Touch the "xxxxV" icon to set the test DC voltage in the
insulation measurement. The screen to the side
appears on the display.
Move the slide bar reference to the desired value for
test voltage by choosing among
50, 100, 250, 500,
1000VDC
Confirm the choice by going back to the initial
measurement screen. Note the presence of the set limit
value.
CAUTION
Disconnect any cable not strictly involved in measurement and especially
check that no cable is connected to In1 input.
Before connecting the test leads, make sure that there is no voltage at the
ends of the conductors to be tested.
5. Connect the alligator clips and/or test leads and/or remote lead to the ends of the
conductors to be tested as in Fig. 10 and Fig. 11