Gx
Series
EN - 95
9. In case of
positive
result of the three test sequentially
performed as
NoTrip
(see § 13.8),
RCD
(see §
13.4) and
M
L/N-PE
the
symbol is shown and the
screen to the side is displayed by the instrument.
Press the
SAVE
button or touch the
icon to save
the measurement (see § 7.1)
10 In case of
negative
result of the
NoTrip
, the auto test
is automatically blocked, the
symbol is shown and
the screen to the side is displayed. Note red value of
the contact voltage
Press the
SAVE
button or touch the
icon to save
the measurement (see § 7.1)
11 In case of
negative
result of the
RCDX1
, the auto test
is automatically blocked, the
symbol is shown and
the screen to the side is displayed. Note red value of
the tripping time
Press the
SAVE
button or touch the
icon to save
the measurement (see § 7.1)
12 In case of
negative
result of the
RCD
, the auto test is
automatically blocked, the
symbol is shown and
the screen to the side is displayed. Note red value of
the tripping current
Press the
SAVE
button or touch the
icon to save
the measurement (see § 7.1)
13 In case of
negative
result of the
M
L-PE
, the auto test
is automatically blocked, the
symbol is shown and
the screen to the side is displayed. Note red value
(lower than minimum set trhreshold) of the L-PE
insulation resistance
Press the
SAVE
button or touch the
icon to save
the measurement (see § 7.1)