Measurement
36
When “CAL” is set in “1.tyPE”
When “CAL” is
set in “1.CAL.t”
Screen for issues an
output in a hold
state during and
after zero calibration
Display only
-
When “trg” is
set
Screen for outputs
the contact for 5
seconds starting 2
seconds after the
completion of
calibration
-
When “CAL.E”
is set
Screen for issues an
output when a
calibration error
occurs
-
Cleaning operation check screen
Cleaning operation check screen
When “no” is
set in “CLn”
Screen for indicating
that the automatic
cleaner is not used
Display only
Note:
“_ _ _ _” indicates that the automatic
cleaner is not used.
-
When “yES” is set in “CLn”
When “nor”,
“or”, or “And” is
set in “tyPE”
Cleaning operation
wait time screen
Displays the time until the next cleaning is
started in a format of hours-minutes-
seconds.
When the automatic cleaner is used, the
countdown of the cleaning cycle timer can
be checked.
Note:
If the time is 10 hours or longer, it is shown
in a format of hours-minutes.
-
When “trg” is
set
Cleaning operation
type
Indicates that “Cleaning operation type
setting” is set to “trg”.
-
-
Cleaning test motion
screen
When ENT is held on, the automatic cleaner
performs one stroke of the reciprocating
motion and stops at the top dead point.
Note:
For test operation
A transmission signal is maintained in a
hold state for the hold time during and after
cleaning.
-
Displays
Display
condition
Item
Description
Setting range