![horiba HU-200TB-W SS-120-W Скачать руководство пользователя страница 86](http://html1.mh-extra.com/html/horiba/hu-200tb-w-ss-120-w/hu-200tb-w-ss-120-w_instruction-manual_2162226086.webp)
Data
74
Option
Cleaner
Note
Each cleaning time varies depending on the pressure of the sample water. The cleaning time is
approximately 30 seconds under open-air conditions on the outlet side while no pressure is applied to
the sample water.
Built-in stand for automatic zero calibration
Product name
Automatic cleaner
Model
SS-AW
Cleaning method
Electric wiper
Power supply
12 V DC, 4 W supplied from HU-200TB-W converter
Cleaning operation
Piston operation is repeated during the cleaning time.
The automatic cleaner waits at the uppermost point when the cleaning time has elapsed.
Cleaning command Cleaning is performed according to the commands sent from the converter via communication.
Mass
Approx. 2.5 kg
Product name Built-in stand for automatic zero calibration
Sample water
condition
Temperature 0
C to 45
C (no freezing)
Pressure
0.02 MPa to 0.3 MPa
Zero water
condition
Temperature 0
C to 45
C (no freezing)
Pressure
Zero filter
Provided: 0.1 MPa to 0.75 MPa
Not provided: 0.02 MPa to 0.75 MPa
Overflow tank
Provided
Overflow tank is provided for the sample water line for removing air bubbles
Material of liquid contact section PVC
Not provided In-line measurement is performed for the sample water without removing air babbles
Zero
calibration unit
Automatic
Sample water, zero water
Automatically switch sample water and zero water
and supply it to a detector
Material of liquid contact section
SCS14, SUS316, FKM, PTFE, SUS303, SUS304,
NBR, SCS13, AS, PP, NBR
Power supply
100 V AC
0
C 50/60 Hz
Manual
Sample water, zero water
Manually switch sample water and zero water and
supply it to a detector
Material of liquid contact section PVC, EPDM, PPF, AS, PP, NBR
Not attached Save zero water in a detector cell and calibration.
: The actuator power supply of the zero calibration unit for 100 V AC specifications, power supply source to this
product is only 100 V AC.