Maintenance
66
Replacing the motor of the automatic cleaner
Replace the drive motor of the automatic cleaner about once every 3 years.
Replacing procedure
1.
Loosen the 4 screws, and remove the cleaner cover.
2.
Check the O-ring for deterioration (elasticity and cracking).
3.
Remove the connector of the wiring of the motor. Do not strain the connector.
4.
Loosen the 4 screws securing the motor platform, and remove the motor assembly.
5.
Take out the set of thrust bearings (retain it for reuse).
6.
Attach a new motor assembly in the reverse order.
7.
Attach the connector of the motor.
8.
Perform a test operation of the automatic cleaner to check the movement (refer to “
Cleaning test motion screen ” (page 36)).
9.
Check all the screws in each section for looseness.
10.
Wipe off any dirt on the O-ring, place the cover on the O-ring, then tighten the
screws.
Note
When screwing a stainless steel bolt into a resin female screw, turn the bolt with your fingers without
using a tool. Screwing in the bolt using excessive torque will damage the screw.
WARNING
Screws
Cleaner cover
Motor
Connector
Thrust bearing
Motor assembly
Motor stand