Measurement
40
Display setting
Selection of
automatic
decimal point
shift or fixed
decimal point
Select either the fixed decimal point or automatic decimal point shift.
Auto: Position of the decimal point is automatically changed according
to the turbidity value.
Fixed decimal point: Select the optimal position of the decimal point
according to the turbidity of the sample water.
10.00: Displays a second decimal place
100.0: Displays a first decimal place
1000 : Displays an integer
Auto,
10.00,
100.0,
1000
Sub display
area display
setting
Specify what is to be displayed on the sub display area when the
measurement mode starts after the end of the hold mode.
non: No display
r1: Displays an alarm value or assignment for relay 1
r2: Displays an alarm value or assignment for relay 2
r3: Displays an alarm value or assignment for relay 3
r1, r2, r3,
non
Setting of a
hidden area
outside of
measurement
range (range
cut)
Specify whether or not to display the area outside the measurement
range when a value exceeds the display range.
yES: Enables the range cut
For fixed decimal point: When a value exceeds the maximum value
of each display range, the value is fixed at the maximum value and
starts blinking.
Fixed decimal point range: The value is fixed to 10.00, 100.0 or
1000.
For the automatic decimal point shift: Refer to the following table.
no: Disables the range cut
For fixed decimal point: When a value exceeds the maximum
value plus 10
of each display range, it is fixed to the value and
starts blinking.
Fixed decimal point range: The value is fixed to 2200, 110.0 or
11.00.
For the automatic decimal point shift: Refer to the following table.
yES, no
Display
Item
Description
Setting
range
yES:
Enables the
range cut
no: Disables
the range
cut
SiO
10.00, 100.0, 1000
10.00,
100.0, 1000
11.00,
110.0, 1100
Auto
0 to 1000
0 to 1100
For
10.00, 100.0, 1000
10.00,
100.0, 2000
11.00,
110.0, 2200
Auto
0 to 2000
0 to 2200
PSL
10.00, 100.0, 1000
10.00,
100.0, 100
11.00,
110.0, 110
Auto
0 to 100
0 to 110.0
Содержание HU-200TB-W SS-120-W