Data
71
Contact output
R3
Contact function
Selectable from among the upper limit alarm, lower limit
alarm, during transmission output hold, cleaning output and
automatic calibration (closed at alarm operation, open under
normal conditions and in a power-off state)
Description of alarm
Setting range: Within turbidity measurement range
Delay time: 0 to 600 (s)
FA
IL
Contact type
Relay contact, SPDT (1c)
Contact capacity
250 V AC 3 A, 30 V DC 3 A (resistance load)
Contact function
Error alarm (closed under normal condition, open in an error or power-off state)
Description of alarm
Error alarms for values outside the measurement range,
self-checks, and calibration errors can be set
Delay time: 0 to 600 (s)
RAN
G
1,
RANG2
Contact type
Relay contact, SPST (1a)
Contact capacity
30 V DC 1 A (resistance load)
Contact function
Status output of the transmission output range
Control output for
automatic zero
calibration (ZERO
OUT, SAMPLE OUT)
Number of output points
2
Output type
Voltage contact output (connection power supply voltage
output)
Contact capacity
250 V AC 3 A/30 V DC 3 A (Resistance load)
Contact function
SAMPLE OUT: Electric ball valve control output
ZERO OUT: Solenoid valve control output
Contact input
Number of input points
4
Contact type
Open collector no-voltage “a” contact
Condition
ON resistance: 100
max.
Open-circuit voltage: 24 V DC
Short-circuit current: 12 mA DC max.
Contact function
EXT1
Switchable between automatic zero calibration command
and transmission hold
EXT2
Switchable between cleaning command and transmission
hold
EXT3, EXT4
Selection command from four ranges using 2-bit contacts
Communication
function
Type
RS-485
Signal form
Two wire type, the power source of the communication is isolated from the
power of measurement circuit (the power source of transmission and
communication are not isolated).
Calibration
Calibration
method
Zero calibration
By filtered clean water
Automatic zero
calibration
Zero calibration is performed automatically by switching to
the filtered clean water (option)
Span calibration Turbidity adjustment method by coefficient input
Compatible standard substances
Kaolin, Formazin (NTU), PSL
Automatic zero
calibration (option)
Zero calibration
method
The electric valve of a sample line is closed, the electromagnetic valve of a
zero line is opened, measured value is automatically judged to be zero within a
time, and zero calibration is carried out.
Setting
Calibration cycle 1 to 999 (h)
Stability time
1 to 60 (s)
Stability level
0.01 degrees/NTU to 0.05 degrees/NTU
Hold time
60 s to 600 s
Cleaning function
(option)
Cleaning method
Electric wiper type (executes cleaning operation by
communicating with the converter)