Maintenance
65
Replacing the wiper of the automatic cleaner
The wiper of the automatic cleaner is made of silicone rubber. As the wiper will gradually
change its size if the number of cleaning times increases, replace it as necessary.
Replace it about once a year for the automatic cleaner.
Replacing procedure
1.
Stop the sample water and remove the automatic cleaner.
2.
Loosen the setscrew securing the wiper support.
3.
Turn the wiper support counterclockwise and replace the wiper.
4.
Attach a new wiper, then turn the wiper support until it stops.
5.
Lightly tighten the setscrew to prevent the wiper support from coming loose.
Note
The deterioration of a wiper varies depending on the sampling water. Replace it as necessary.
Do not tighten the setscrew too tightly.
WARNING
Allen wrench
Setscrew
Wiper support
Wiper