Measurement
57
List of items displayed for the user check menu
Display
Item
Description
LCD check
LCD check screen
The whole LCD lights up to find defects of the display panel.
Relay output check: R1 to R3
Relay output 1 check
screen
The ON/OFF status for relay 1 (r1) can be checked.
ENT: ON/OF switching
Note:
When the system enters this mode, the contact output is set to OFF.
Relay output 2 check
screen
The ON/OFF status for relay 2 (r2) can be checked.
ENT: ON/OF switching
Note:
When the system enters this mode, the contact output is set to OFF.
Relay output 3 check
screen
The ON/OFF status for relay 3 (r3) can be checked.
ENT: ON/OF switching
Note:
When the system enters this mode, the contact output is set to OFF.
FAIL output check
screen
The ON/OFF status of FAIL can be checked.
(When a FAIL occurs, the coil current of the relay is cut off. The current
flows through the coil of the relay in a normal state.)
Range contact output
1 check screen
The ON/OFF status for range contact output 1 can be checked.
Range contact output
2 check screen
The ON/OFF status for range contact output 2 can be checked.
Screen for checking
electric ball valve
output for sample
water line
The electric ball valve operation can be checked.
on: The valve is closed
oFF: The valve is open
Screen for checking
solenoid valve output
for zero water line
The solenoid valve operation can be checked.
Note:
When the output is “on”, voltage is output and the valve is opened.