Installation
19
Power supply
A power switch is equipped beside the terminal block.
The voltage of AC power source is free between 100 V and 240 V AC.
The terminal screws of the contact output are M4 type.
The applicable electric wire is of 0.75 mm
2
to 5.5 mm
2
(AWG18 to AWG10).
Provide a power switch near the HU-200TB-W so that the power can be turned ON/OFF.
Note
The operation of this equipment at any voltage outside the rated range may cause damage.
It is recommended that the fluctuations in power supply voltage fall within a range of
10
.
As illustrated below, connect the power cable and the grounding cable.
Instructions for grounding
Separate this grounding from any other grounding for electric equipment such as the motor.
WARNING
Be sure to check that no electric power is supplied before starting the procedure.
WARNING
Be sure to ground the grounding terminal (The ground resistance should be 100
Power switch
ON
OFF
Power supplied
Voltage: 100 V to 240 V AC
Frequency: 50 Hz/60 Hz
(L) (N)
The ground resistance should be 100
or less