Measurement
46
Cleaning
operation type
setting
Specify the way to start cleaning with a internal timer or a combination of
external command.
nor: Cleaning will start in accordance with the internal timer setting.
The external command can be used to hold the transmission output.
or: Cleaning is performed in accordance with the internal timer or
when an external command is input.
And: Cleaning is performed in accordance with the internal timer while
the external command is input.
trg: Cleaning will start only by the external signal triggered at least 2
seconds. The internal timer setting is neglected.
Refer to “ Cleaning start condition ” (page 51).
Note:
This is displayed when “CLn” is set to “yES”.
nor, or,
And, trg
Internal timer
cleaning cycle
setting
Specify the cleaning cycle in hours.
Note:
This is displayed when “CLn” is set to “yES”, and “tyPE” of CLn is set to
“nor”, “or”, or “And”.
0.1 to
168.0 (h)
Cleaning time
setting
Specify the cleaning time in seconds.
Note:
This is displayed when “CLn” is set to “yES”.
20 to 600
(s)
Setting of
retention time
for
transmission
output hold
after
completion of
cleaning
Specify the hold retention time in seconds after the completion of the
cleaning for the specified extended time.
Note:
When “CLn” is set to “no”, the hold retention time is effective for the test
cleaning.
60 to 600
(s)
External input 1 (calibration/hold)
Selection of
command by
contact
(calibration/
hold command)
When external input 1 is a calibration command (“Setting of automatic
zero calibration start method” is set to “or”, “And”, or “trg”), if this contact
is closed, zero calibration and its associated operations will be started
automatically.
Note:
The following display cannot be changed.
When “tyPE” of Automatic zero calibration setting is set to “nor”
(internal timer only), “HoLd” is displayed
When “tyPE” of Automatic zero calibration setting is set to “or”, “And”,
or “trg”, “CAL” is displayed
CAL,
HoLd
Setting for
reversing logic
of hold
command
(calibration/
hold command)
When “tyPE” of Automatic zero calibration setting is set to “nor”, the logic
of the calibration/hold command is reversed.
no: Contact is closed when a calibration command is issued
yES: Contact is opened when a hold command is issued
yES, no
Display
Item
Description
Setting
range