Safety Functions
31
en
7
Safety Functions
The electronic components continuously monitor the proper function of the sys-
tem. System malfunctions are indicated with audible warning beeps, error mes-
sages, and/or the blocking of system functions. A table listing a summary of
possible error and warning messages is provided in Chapter 10, Error and Warn-
ing Messages.
Intrauterine pressure 10 mmHg above pre-
set intrauterine pressure setting
If the intrauterine pressure exceeds the intrauterine pressure setting by
10 mmHg for longer than 5 seconds, the pressure reduction function is activated.
The roller wheel will move forward or backward a few times during the pressure
reduction process. If the pressure cannot be reduced, the message
Overpressure/
Open Stopcock
is displayed and three warning beeps are emitted.
Intrauterine pressure > 150 mmHg
The message
Maximum Pressure
is displayed and the pump will emit 3 warning
beeps once the intrauterine pressure exceeds 150 mmHg. The maximum permis-
sible pressure has now been reached.
Intrauterine pressure > 200 mmHg
If the intrauterine pressure exceeds 200 mmHg for longer than 5 seconds, the
roller wheel stops and the message
Overpressure/Check Stopcock
is displayed.
Three short continuous audible warning beeps are emitted until the pressure is
reduced. Once the intrauterine pressure falls below 200 mmHg, the audible
warning beeps stop and the pump wheel resumes turning automatically.
Check Tube Set Installation
If the inflow tube set is not inserted properly into the roller wheel, pressing the
Prime
button results in a short audible beep and
Check Tube Set Installation
is
displayed. The roller wheel does not start to turn.
Pressure measuring system malfunctions
If a malfunction is detected in the pressure measurement electronics,
Sensor Er-
ror
is displayed and five short warning beeps are emitted. The roller wheel stops
turning.
Scale overload
If the maximum permissible weight of the scale is exceeded (bag scale or con-
tainer scale, a continuous audible warning beep is emitted and
Scale Overloaded
Check Scale
is displayed. The warning stops once the excess weight is removed
from the scale.
Loading/unloading scale while in operation
The following messages are displayed when a container is removed from the
scale or a new fluid bag is added during operation of the system:
Container
Change, Press Resume or Bag Change, Please Proceed
.
Pressure setting at restart
If the last pre-set intrauterine pressure setting is greater than 80 mmHg, this val-
ue is reset to the default value of 80 mmHg.
Deficit limit
Each additional deficit increase by 100 ml above the selected deficit threshold
triggers 3 warning beeps that are repeated while the pump continues to operate.
The message
Deficit Limit Exceeded
is displayed.
Deficit rate >300 ml/min
If the deficit rate exceeds 300 ml/min, three audible warning beeps are emitted
and the message
High Fluid Loss Check Leakage
is displayed. If no obvious source
of high fluid loss can be identified, an assessment of potential cervical or uterine
perforation should be made.
Serious system defect
Five short audible warning beeps are emitted and the message
Motor Error
is dis-
played. It is possible that errors can occur in the start up sequence prior to the en-
abling of the Pump Display. In this situation, the Pump Display will remain blank.
Содержание Aquilex
Страница 4: ......
Страница 8: ......
Страница 10: ......
Страница 61: ...Appendix 53 en 16 Appendix 16 1 Test Log Date Tests Performed Results Comment Signature ...
Страница 64: ......
Страница 118: ...56 Anhang de 16 Anhang 16 1 Testprotokoll Datum Durchgeführte Tests Ergebnis Bemerkung Unterschrift ...
Страница 120: ......
Страница 122: ......
Страница 176: ...56 Annexe fr 16 Annexe 16 1 Procès verbal de tests Date Tests réalisés Résultat Remarque Signature ...
Страница 178: ......
Страница 180: ......
Страница 234: ...56 Apéndice es 16 Apéndice 16 1 Registro de test Fecha Test realizados Resultado Anotaciones Firma ...
Страница 236: ......
Страница 238: ......
Страница 292: ...56 Appendice it 16 Appendice 16 1 Protocollo di verifica Data Test eseguiti Risultato Nota Firma ...
Страница 294: ......
Страница 296: ......
Страница 351: ...Bijlage 57 nl 16 Bijlage 16 1 Testprotocol Datum Uitgevoerde tests Resultaat Opmerking Handtekening ...
Страница 353: ......
Страница 354: ......