Error and Warning Messages
en
10
Error and Warning Messages
The messages appear in the pump display and warning beeps are heard. Beeps
indicating a warning or operating message are sounded several times.
Message in the pump display
Warning beeps
Method
Check Tube Set Installation
1 warning beep
Remove and re-insert tube set. If the
message reappears, insert a new tube
set.
Tube Set Over Usage Limits
1 warning beep
Tube set detection indicates that the
tube set has been used already. Insert
new tube set.
Check Flow Path, Stopcock, Clamps
3 warning beeps
Flow path is blocked. Check that bag
clamps and hysteroscope stopcock are
open. Make sure the tube set is not
blocked.
Incorrect Tube Set
1 warning beep
Replace tube set. The tube set does not
match the type approved for the Aquilex
System.
Pump Paused, Press Resume
1 warning beep
Pause/Resume button has been acti-
vated. Press Pause/Resume button again
to resume surgery.
Overpressure Open Stopcock
3 warning beeps
Most commonly triggered when hystero-
scope stopcock is closed while pump is
operating at peak flow rate. Open hys-
teroscope stopcock or remove other clo-
sure to relieve pressure.
Overpressure Check Stopcock
3 beeps, constantly repeating until pres-
sure is reduced
Pressure has exceeded the 200 mmHg
safety limit and must be reduced. The
most common cause is a closed hystero-
scope stopcock while the pump operates
at the highest flow rate.
Open inflow stopcock on hysteroscope or
remove blockage clamping off the inflow
tube set.
Maximum Pressure
No beep
The message "Maximum Pressure" is
depicted when the intrauterine pressure
is greater than 150 mmHg.
Deficit Limit Reached
3 warning beeps
Actions are at the discretion of the physi-
cian.
Deficit Limit Exceeded
3 warning beeps
Actions are at the discretion of the physi-
cian. If necessary, perform manual defi-
cit control.
Pressure Threshold
1 warning beep
For a hysteroscopy, pressures above
100 mmHg are usually not necessary.
Careful monitoring of the fluid deficit is
recommended.
Check Scale Connection
3 warning beeps
Check scale connection. Reconnect scale,
restart device. If the message reappears,
contact Hologic.
Remove Tube Set for System Check
1 warning beep
Make sure all tube sets are removed
from the roller wheel during the system
test. Remove tube set and wait until you
hear a warning beep and the message
"Insert Tube Set" appears.
Prime Fail-Open Stopcock, Clamps
3 warning beeps
Check bag clamp(s) and hysteroscope
inflow stopcock are open. Press the
"Prime" button to restart.
High Fluid Loss Check Leakage
3 warning beeps
Actions are at the discretion of the physi-
cian. If necessary, perform manual defi-
cit control.
Содержание Aquilex
Страница 4: ......
Страница 8: ......
Страница 10: ......
Страница 61: ...Appendix 53 en 16 Appendix 16 1 Test Log Date Tests Performed Results Comment Signature ...
Страница 64: ......
Страница 118: ...56 Anhang de 16 Anhang 16 1 Testprotokoll Datum Durchgeführte Tests Ergebnis Bemerkung Unterschrift ...
Страница 120: ......
Страница 122: ......
Страница 176: ...56 Annexe fr 16 Annexe 16 1 Procès verbal de tests Date Tests réalisés Résultat Remarque Signature ...
Страница 178: ......
Страница 180: ......
Страница 234: ...56 Apéndice es 16 Apéndice 16 1 Registro de test Fecha Test realizados Resultado Anotaciones Firma ...
Страница 236: ......
Страница 238: ......
Страница 292: ...56 Appendice it 16 Appendice 16 1 Protocollo di verifica Data Test eseguiti Risultato Nota Firma ...
Страница 294: ......
Страница 296: ......
Страница 351: ...Bijlage 57 nl 16 Bijlage 16 1 Testprotocol Datum Uitgevoerde tests Resultaat Opmerking Handtekening ...
Страница 353: ......
Страница 354: ......