Qty Location
Type
Size/length
Torque
Bit type
1
Battery door
Shoulder screw
M3 x 0.5
3.0 in-lb ±0.5 in-lb #1 Phillips
2
Rear housing
Pan head machine
screw
M3 x 0.5
6.0 in-lb ±0.5 in-lb #1 Phillips
4
PCBA mount
Plastite
#4 x 0.25
4.0 in-lb ±0.5 in-lb Torx T10
Connectors
Connector types
Disassembly and repair procedures require that you disconnect and reconnect the following
connector types:
•
Locking (squeeze-release)
: Locking connectors use a latching mechanism to prevent
accidental disconnection during assembly and use. The latch is located on one end of a tab so
it may flex and lock into place when coupled with its matching connector. The tab provides a
lever to release the latch. When disconnecting, squeeze to provide pressure on the tab to
unlatch. Some connectors have multiple latches that require you to press multiple tabs to
release.
To remove a locking connector, squeeze the release lever and remove the cable.
To connect a locking connector, push the mating pieces together until the latch locks in place.
•
Pressure
: Pressure connectors use friction to prevent accidental disconnects.
To remove a pressure connector, grasp each connector mating half and pull the halves apart.
CAUTION
Do not use excessive force to disconnect the
connector. Excessive force may result in pulling the mounted
connector off the circuit board.
To connect a pressure connector, grasp each connector mating half and insert one half into
the other.
•
ZIF (zero insertion force)
: ZIF connectors use a sliding outer piece that latches and unlatches
to secure and release the flex cable. ZIF cables cannot be successfully connected or
disconnected without properly unlatching and latching the sliding outer piece.
CAUTION
Do not use excessive force when releasing pressure on
the connector. Excessive force may result in breaking the sliding
outer piece.
To remove a ZIF connector
CAUTION
Remove a flex cable only
after
the ZIF latch is open.
1. Using a suitable tool (for example, a spudger or needle-nose pliers), slide the latching
piece of the connector away from the connector body.
2. Remove the cable.
Service Manual
Appendices 57
Содержание Welch Allyn Connex ProBP 3400
Страница 1: ...Welch Allyn Connex ProBP 3400 digital blood pressure device Service Manual Software version 1 04 XX ...
Страница 8: ...4 Symbols Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 12: ...8 Safety Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 20: ...16 Overview Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 22: ...18 Controls and indicators Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 24: ...20 Connections Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 26: ...22 Power up sequence Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 42: ...38 Disassembly and repair Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 46: ...42 Reassembly Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 64: ......