Reassembly
Device reassembly can be performed by reversing the disassembly instructions. However, the
following alternate procedure simplifies device reassembly after the chassis assembly and circuit
board assembly have been separated.
Before you begin this procedure:
•
Reassemble the chassis assembly and reassemble the circuit board assembly by reversing the
disassembly instructions. Keep the assemblies separated.
•
Ensure that the isolation bumpers are installed on the side panels as described in “When
replacing the side panels or isolation bumpers.”
1. Orient the chassis and circuit board assemblies as shown below, and connect the dump valve
cable to J10 on the main board.
2. Place the front housing face down on the antistatic mat, and insert one side panel.
3. Place the circuit board assembly into the front housing.
a. Turn over the chassis and circuit board assemblies.
b. Lower the circuit board assembly into the front housing.
39
Содержание Welch Allyn Connex ProBP 3400
Страница 1: ...Welch Allyn Connex ProBP 3400 digital blood pressure device Service Manual Software version 1 04 XX ...
Страница 8: ...4 Symbols Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 12: ...8 Safety Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 20: ...16 Overview Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 22: ...18 Controls and indicators Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 24: ...20 Connections Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 26: ...22 Power up sequence Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 42: ...38 Disassembly and repair Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 46: ...42 Reassembly Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 64: ......