Service record
Date
:
Time
:
Device name
:
Serial number
:
Technician
:
MODF firmware version
:
Host Controller version
:
Leak test
Specification
Actual reading
LeakTest:
Max: 5
Overpressure test
Specification
Actual reading
OverPressureTest:
280 . . . . 329
Accuracy test
Target pressure ±5
mmHg
Pressure meter
Service tool
Specification
Pass/fail
0 mmHg
mmHg
mmHg
±3 mmHg
50 mmHg
mmHg
mmHg
±3 mmHg
150 mmHg
mmHg
mmHg
±3 mmHg
250 mmHg
mmHg
mmHg
±3 mmHg
46 Functional verification tests