3. Disconnect the dump valve cable from J10 on the main board.
4. Disconnect the dump valve manifold from the pressure transducer on the blood pressure
board.
5. Lift the chassis assembly away from the circuit board assembly, and disconnect the battery
contact harness from J2 on the main board.
Reassembly notes
See the “Reassembly” section for the recommended procedure to reassemble the device after the
chassis and circuit board assemblies have been separated.
32 Disassembly and repair
Welch Allyn® Connex® ProBP™ 3400 digital blood pressure device