c. Align the pegs on the PCBA mount with the isolation bumpers on the side panel.
d. Ensure that the flex cable is not under the circuit board assembly.
4. Insert the second side panel into the front housing, aligning the isolation bumpers with the
PCBA mount pegs.
5. Insert the flex cable into the ZIF connector (J5) on the main board and lock the connector.
6. Pivot the chassis assembly over the circuit board assembly, and connect the battery contact
harness to J2 on the main board.
7. Connect the step valve manifold to the step valve on the blood pressure board.
8. Connect the pump cable to J80 on the blood pressure board. Ensure that the cable is routed
as shown below.
40 Reassembly
Welch Allyn® Connex® ProBP™ 3400 digital blood pressure device
Содержание Welch Allyn Connex ProBP 3400
Страница 1: ...Welch Allyn Connex ProBP 3400 digital blood pressure device Service Manual Software version 1 04 XX ...
Страница 8: ...4 Symbols Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 12: ...8 Safety Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 20: ...16 Overview Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 22: ...18 Controls and indicators Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 24: ...20 Connections Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 26: ...22 Power up sequence Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 42: ...38 Disassembly and repair Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 46: ...42 Reassembly Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 64: ......