No.
Item
No.
Item
16
Dump valve
20
Pump manifold
17
Dump valve manifold
21
Blood pressure port
18
Dump valve hose
22
Step valve hose
19
Air filter
23
Step valve manifold
Remove the blood pressure port
Before you begin, remove the following as described in this manual:
•
Battery
•
Rear housing
1. Lift the blood pressure port to disconnect it from the inner chassis.
2. Disconnect the blood pressure port from the pump manifold.
Remove the pump and manifold assembly
Before you begin, remove the following as described in this manual:
•
Battery
•
Rear housing
•
Front housing and side panels
•
Circuit board assembly
1. On the underside of the inner chassis, do the following:
a. Remove the pump cable from the routing slot and wire clip.
b. Remove the air filter from the retaining clip.
2. Disconnect the dump valve manifold from the dump valve.
3. Remove the step valve and dump valve hoses from the hose clips on the inner chassis.
34 Disassembly and repair
Welch Allyn® Connex® ProBP™ 3400 digital blood pressure device
Содержание Welch Allyn Connex ProBP 3400
Страница 1: ...Welch Allyn Connex ProBP 3400 digital blood pressure device Service Manual Software version 1 04 XX ...
Страница 8: ...4 Symbols Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 12: ...8 Safety Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 20: ...16 Overview Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 22: ...18 Controls and indicators Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 24: ...20 Connections Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 26: ...22 Power up sequence Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 42: ...38 Disassembly and repair Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 46: ...42 Reassembly Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 64: ......