No.
Item
No.
Item
2
Battery compartment door
7
Isolation bumper
3
Battery
8
Chassis and circuit board assemblies
4
Screw, M3 x 0.5 pan head
9
Front housing with key panel
5
Rear housing
Remove the battery
Before you begin, ensure that the device is powered off.
1. Loosen the screw to remove the battery compartment door.
2. Remove the battery.
Remove the rear housing
Before you begin, remove the battery as described in this manual.
1. Remove the two rear-housing screws.
2. Lift the rear housing on the side closest to the bottom of the device and pivot it toward the
top of the device until it comes free of the front housing.
Reassembly notes
•
Start by orienting the rear housing at an angle of approximately 45 degrees from the front
housing and then mate the tabs on the rear housing with the slots on the front housing.
•
Verify that no wires or hoses are crushed or pinched before securing the rear housing with
screws.
•
After tightening the screws, examine the seams between the rear housing, side panels, and
front housing to verify that there is no bowing. Bowing indicates that something is preventing
the mating surfaces from seating properly.
Remove the front housing and side panels
Before you begin, remove the following as described in this manual:
Service Manual
Disassembly and repair 29
Содержание Welch Allyn Connex ProBP 3400
Страница 1: ...Welch Allyn Connex ProBP 3400 digital blood pressure device Service Manual Software version 1 04 XX ...
Страница 8: ...4 Symbols Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 12: ...8 Safety Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 20: ...16 Overview Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 22: ...18 Controls and indicators Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 24: ...20 Connections Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 26: ...22 Power up sequence Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 42: ...38 Disassembly and repair Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 46: ...42 Reassembly Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 64: ......