Disassembly and repair
These procedures provide instructions for device disassembly and board removal. Except where
otherwise noted, the assembly procedure is the reverse of the disassembly procedure.
An exploded view of the assembly precedes disassembly instructions, with callouts referencing the
parts.
Each part’s disassembly instructions might include one or both of the following:
•
Reassembly notes
: This contains information specific to reassembly not addressed in the
disassembly instructions.
•
When replacing the
component
: This contains information specific to installing a
replacement part.
For information about screws or connectors used in the device, see “Screws” and “Connectors” in
the appendices.
NOTE
After performing any of these procedures and before returning the device to
service, you must use the service tool, Gold edition, to complete the full suite of
functional tests to ensure that all systems are operating within the design
specifications. For more information about these tests and the service tool, see
“Functional verification and calibration.”
If you do not have the service tool, contact Welch Allyn Technical Support.
WARNING
Electrical shock hazard. Disconnect AC power before opening the
device. Disconnect and remove the battery before proceeding with disassembly.
Failure to do this can cause serious personal injury and damage to the device.
WARNING
Risk of fire, explosion and burns. Do not short-circuit, crush, incinerate,
or disassemble the battery.
WARNING
Safety risk. Do not attempt to service the device when the device is
connected to a patient.
CAUTION
Before disassembling the device, disconnect the AC power cord and
any attached accessories.
CAUTION
Perform all repair procedures at a static-protected station.
CAUTION
When the device case is opened, regard all parts as extremely fragile.
Execute all procedure steps with care and precision.
27
Содержание Welch Allyn Connex ProBP 3400
Страница 1: ...Welch Allyn Connex ProBP 3400 digital blood pressure device Service Manual Software version 1 04 XX ...
Страница 8: ...4 Symbols Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 12: ...8 Safety Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 20: ...16 Overview Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 22: ...18 Controls and indicators Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 24: ...20 Connections Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 26: ...22 Power up sequence Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 42: ...38 Disassembly and repair Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 46: ...42 Reassembly Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 64: ......