Appendices
Clean the device
WARNING
Electric shock hazard. Before cleaning the device, disconnect the
power cord from the power source and the device. Ensure power transformer and
wall plug assembly are dry before plugging into an outlet.
WARNING
Take care to prevent water or other fluid from entering any connectors
on the device, power transformer, and wall plug. Should this occur, dry the
connectors with warm air. Check the accuracy of all operating functions.
CAUTION
The device is not heat-resistant. Do not autoclave.
The following agents are compatible with the device:
•
70 percent isopropyl alcohol
•
10 percent chlorine bleach/90 percent water solution
NOTE
Clean the device on a routine basis, according to your facility's protocols
and standards or local regulations.
70 percent isopropyl alcohol
Wipe the device with a clean cloth slightly dampened with 70 percent isopropyl alcohol.
55
Содержание Welch Allyn Connex ProBP 3400
Страница 1: ...Welch Allyn Connex ProBP 3400 digital blood pressure device Service Manual Software version 1 04 XX ...
Страница 8: ...4 Symbols Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 12: ...8 Safety Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 20: ...16 Overview Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 22: ...18 Controls and indicators Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 24: ...20 Connections Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 26: ...22 Power up sequence Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 42: ...38 Disassembly and repair Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 46: ...42 Reassembly Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 64: ......