Battery
•
Rear housing
1. Unlock the ZIF connector (J5) on the main board by sliding the latching piece of the connector
away from the connector body.
CAUTION
Do not use excessive force when releasing pressure on
the connector. Excessive force may result in breaking the sliding
outer piece.
2. Remove the flex cable from the ZIF connector.
3. Grasp both side panels and lift them and the chassis and circuit board assemblies away from
the front housing.
4. While holding the chassis and circuit board assemblies, remove the side panels.
CAUTION
Always protect the LCD. When the main board is outside of the device
case, never rest the board on the LCD.
When replacing the front housing
Remove the liner from the inside of the housing window if present. Wipe any dust or fingerprints
off of the window with a clean cloth.
Reassembly notes
Before reassembly, do the following to ensure proper installation of the side panels and isolation
bumpers:
1. Place the front housing and side panels on the antistatic mat as shown in the diagram below,
making sure that the arrows on the side panels point away from the front housing.
30 Disassembly and repair
Welch Allyn® Connex® ProBP™ 3400 digital blood pressure device
Содержание Welch Allyn Connex ProBP 3400
Страница 1: ...Welch Allyn Connex ProBP 3400 digital blood pressure device Service Manual Software version 1 04 XX ...
Страница 8: ...4 Symbols Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 12: ...8 Safety Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 20: ...16 Overview Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 22: ...18 Controls and indicators Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 24: ...20 Connections Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 26: ...22 Power up sequence Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 42: ...38 Disassembly and repair Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 46: ...42 Reassembly Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 64: ......