Doc -030 Hatch comfort C1 User manual
Revised on 15/06/2021
10.2.2
HP lubrication oil particle collection bottle
CAUTION
Check the condition of the bottle located in the bottom of the dentist table
at least monthly.
CAUTION
Oil absorbed by a filter is flammable. Dispose the filter according to the local
regulation.
1)
Hold the bottle and lid. Turn it counterclockwise to disassemble. Throw away oil
particles from the bottle.
2)
Pull out the filter attached on a shaft.
3)
After rinsing out the bottle and filter with running
water, dry them completely.
4)
Insert a new filter over the shaft. Assemble the bottle.
5)
Pick up and operate a handpiece one by one to check
air flow. If not working properly, stop the operation
and contact a service center.
NOTE
If O-rings in the edge of the air turbine connector are damaged, they can
cause air leakage.