Doc -030 Hatch comfort C1 User manual
Revised on 15/06/2021
..
Oil Collector
Air adjustment valve
8.7
Volume adjustment of air/water on instruments
8.7.1
Instruments in dentist table
▶ Adjust volume of each
instrument with a control knob located in the front
lower of the table
.
3-WAY
TURBINE
MOTOR
SCALER
AIR
WATER
AIR
WATER
①
②
③
④
⑤
⑥
⑦
Knob No
Function
Remark
①
Air volume for holder 1
1)
The knob location is
based on the standard
layout of instruments.
The position of a knob
may vary based on a
layout change.
2)
Turn the knob
counterclockwise to
increase the volume.
②
Water volume for holder 1
③
Water volume for holder 2
④
Water volume for holder 3
⑤
Air volume for holder 4
⑥
Water volume for holder 4
⑦
Water volume for holder 5
8.7.2
3-WAY syringe in assistant table
▶
To adjust the air/water volume for a 3-WAY
syringe in the assistant table, open the
panel in the left of the table and
adjust the valve mounted on the right side.
Water adjustment valve
8.7.3
HP oil collector
▶
Located at the bottom of the dentist table.
▶
The container collects water and lubrication oil
particles generated from high speed handpieces.
Check inside monthly and dispose the debris collected