Doc -030 Hatch comfort C1 User manual
Revised on 15/06/2021
knob
Rinsing pipe
▶
Disassembling rinsing pipe
When disassembling the rinsing pipe,
Loosen the knob and pull out the pipe.
Assemble in the reverse order.
10.1.6
Cleaning vacuum system
10.1.7
Cleaning water lines
Rinsing pipe
CAUTION
Clean the water line at least once daily before using it.
This process is important specially if the dental chair unit has not been in
operation for a few days.
▶
Water remained in the warmer tank and lines should be drained out first.
▶
Tumbler filler
1)
Activate the tumbler filler by putting cup.
2)
Throw away water and fill up the cup again. Repeat this 8 times.
3)
If it is required to fill a cup manually, press
‘tumbler
water
supply’
switch in the
control panel for one minute.
CAUTION
When disassembling the rinsing pipe, pay attention not to damage or lose the
rubber ring.
▶
When assembling the rinsing pipe, tighten the know firmly to avoid water
leakage.
CAUTION
Clean the vacuum system at least once daily.
CAUTION
Use only a recommended cleaning detergent (Oratol Plus).
Please follow the cleaning instruction of each detergent.
▶
If unsuitable detergents are used, tubes, lines and other parts may get
damaged and cause water leakage or weakening the suction power.