Doc -030 Hatch comfort C1 User manual
Revised on 15/06/2021
Fuse holder cap
9.2
Fuse for main transformer
A fuse is located inside the chair base.
The fuse sizes are
Φ6㎜
diameter, 30
㎜
length.
Main transformer fuses should be replaced in the
following sequence.
1)
Turn OFF the main switch.
2)
Take off the PCB protection cover by loosening
fixation screws.
3)
Remove the fuse holder cap.
4)
Remove burnt fuses and put new ones.
Fuse (4A / 250V)
5)
Put back the fuse holder cap.
6)
Assemble the PCB protection cover.
PCB protection cover